1. Arrangement of control elements 

.

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

2. General description 

.

 

The instrument includes two common pressure-balances of the basic series CPB5000. 
For the calibration of differential gauges at a static line pressure, the systems can be seperated. 
 
One system is used to generate the line pressure. The second one supplys the differential pressure 
(including line pressure). Following the description for the use of the instrument. 
 
 

3. Working principle 

.

 

 

Piston 2 

Inlet Outlet 

Spindle 

Bypass 

Differential pressure 

Static line pressure 

Outlet 

Inlet 

Piston 1 

Device under test 

Spindle 

 

 

2 connections for 
test specimen 

Initial pressure pump 

Fixture for 
ConTect-System 

Water 
level 

Spindle pump 

Screwed drain plug for 
tank

 (hydraulic only)

 

Fixture for 
ConTect-System 

Inlet valve 

(pneumatic only) 

Outlet 
valve 

2 Test pressure 
gauges 

Bypass-valve 

Inlet valve 

(pneumatic only) 

Outlet 
valve 

Summary of Contents for CPB 5000 DP

Page 1: ...ebsanleitung CPB 5000 DP Addition to the operating instructions of the CPB 5000 Erg nzung zur Betriebsanleitung CPB 5000 Differential Pressure Balance GB Differential Pressure Balance CPB 5000 DP Diff...

Page 2: ...Balance Version 1 0 2 Information This symbol provides you with information notes and tips Warning This symbol warns you against actions that can cause injury to people or damage to the instrument Ope...

Page 3: ...to generate the line pressure The second one supplys the differential pressure including line pressure Following the description for the use of the instrument 3 Working principle Piston 2 Inlet Outlet...

Page 4: ...spindle pump Now both systems should be in floating position If both systems float and spin the device under test is applied to a zero pressure Note but each port has an additional line pressure The s...

Page 5: ...erell dient ein Messsystem zur Erzeugung des statischen Druckes Mit dem zweiten ConTect System erfolgt die Beaufschlagung des Pr flings mit dem Differenzdruck inkl statischen Vordruck Die Vorgehenswei...

Page 6: ...liegt am Pr fling ein Differenz Nulldruck 0 bar an Die Sinkgeschwindigkeiten sollten vergleichbar sein Gegebenenfalls die Auflagen am statischen Kolben mit Feinmassen anpassen 3 2 Einstellen des Diff...

Page 7: ...E mail info wika com my www wika com my USA WIKA Instrument Corporation Lawrenceville GA 30043 Phone 1 770 513 82 00 Fax 1 770 338 51 18 E mail info wika com www wika com Asia China WIKA Internationa...

Page 8: ...16 E mail info wika donetsk ua Bulgaria WIKA Bulgaria EOOD 1309 Sofia Phone 359 2 82138 10 Fax 359 2 82138 13 E Mail t antonov wika bg Russia ZAO WIKA MERA 127015 Moskau Phone 7 495 786 21 25 Fax 7 4...

Reviews: