5. Commissioning, operation
5.6 Setting the response time
The response time affects how fast the transmitter reacts to changes
in the pressure conditions in the system. The response time defines
the time the measuring instrument needs in order to reach 63 % of
the measured value. For unstable pressure conditions, select a longer
response time.
Example:
Selected response time: 4.0 seconds
Result: The output signal reaches a new value in 20 seconds (response
time * 5)
In order to change the response time, install or remove a jumper in slot J4.
■
Jumper in slot J4 - 4.0 seconds response time.
■
No jumper in slot J4 = 0.8 seconds response time