6. Commissioning, operation
Parameters Values
Description
Key 1
Keys 2 and 3
ScAL
Slope (scale)
-5.00 … 5.00
The input of the slope correction factor is made
in %. The displayed value is calculated as per the
following equation:
Display = (measured value - Offset - di.Lo) * (1 +
slope correction [% / 100] ) + di.Lo
Example for offset and slope correction
Connection of a pressure transmitter
The instrument display without offset and slope correction is as follows:
at 0 bar = 0.08, at 20 bar = 20.02
From this is calculated:
Zero point:
0.08
Slope:
20.02 – 0.08 = 19.94
Deviation:
0.06 = set slope - actual slope = 20.00 – 19.94
Therefore the following should be set:
Offset =
0.08 = zero error
Scale =
0.30
= deviation / actual slope = 0.06 /19.94 = 0.0030 = 0.30 %
Min/Max value memory
The instrument has a Min/Max value memory. In this are stored the lowest
and the highest diplayed values. The Min/Max value memory is a volatile
memory, meaning the stored values are lost when the power is interrupted.
Key
Display
To recall Min
value:
Press key 3 briefly
"Lo"
will be displayed briefly and then for
approx. 2 s the Min value will be shown.
To recall Max
value:
Press key 2 briefly
"Hi"
will be displayed briefly and then for
approx. 2 s the Max value will be shown.
To delete
the Min/Max
values:
Press keys 2 and
3 simultaneously
for 2 s.
"CLr"
will briefly be shown in the display,
the Min/Max value will be reset to the
current display value.