Battery replacement
Remove the probes from any testing point, when opening the battery case.
Batteries are empty when the continuity test with both test probes connected
cannot be done anymore. A battery symbol in the LCD indicates low battery.
Follow the procedure below and replace batteries with new ones (type IEC
LR03 1.5 V).
• Unscrew the battery door using Philips type screwdriver.
• Pull out the battery door and replace the batteries. Insert new batteries accor-
ding to the engraving on the battery door.
• Re-assemble battery door.
Confirm that the battery door case is properly locked prior to measurements.
Technical data
Voltage range
0.5 V...1,000 V AC True RMS (16 2/3...950 Hz), 0.5
V...1,500 V DC(±)
LED nominal voltage
12/24/50/120/230/400/690/1,000 V, AC (16 2/3�950 Hz),
DC(±)
LED tolerances
according to EN 61243-3
ELV indication LED
> 50 V AC, > 120 V DC
Response time
< 0.5 s (LED); < 1 s (LCD)
LCD range
0.5 V...1,000 V AC, 1,500 V DC
LCD resolution
0.1 V(< 10 V) and 1 V (> 10 V)
LCD accuracy
3 % + 3 digit > 10 V, 3 % + 5 digit < 10 V
Crest factor
1 ... 10V: CF > 2.5 / >10 ... 300 V: CF=5 / >300V�500V:
CF=3 / >500V�750V: CF=2 / >750V CF=1.5
LCD overrange indication
“OL”
Peak current
Is < 3.5 mA (at 1,000 V)
Measurement duty
30 s ON (operation time), 240 s OFF (revocery time)
Internal battery consumption approx. 80 mA
Single-pole phase test
voltage range
100�1,000 V AC (40...70 Hz)
Phase rotation test
170�1,000 V phase-to-phase, AC 40...70 Hz
Continuity test
detection range 0�500 kΩ + 50 %
NCV test
100...1,000 V AC against earth (50/60 Hz)
Resistance measurement
detection range 0�1,999 Ω ± (5 % + 10 dgt); resolution:
1 Ω
Frequency measurement
1...950 Hz +/- (5 % + 5-digit); resolution: 1 Hz; 10...1,000
V AC
Battery
3 V (IEC LR03 1.5 V x 2)
Diode test
0...2 V (+/- 5 % 10-digit); resolution: 0.01 V
Temperature
-15�50 °C operation; -20�60 °C storange, No
condensation
Humidity
max. 85 % RH
Altitude
up to 2,000 m
Overvoltage
CAT. III 1,000 V / CAT. IV 600 V
Standard
EN 61243-3:2014
Pollution degree
2
Protection
IP64