Thermometer Calibration Instructions 

1.  Fill the tub so that the bottom of the thermometer tip is 

immersed in at least 10.2 cm (4 inches) of water.  

2.  Place the stem tip of a calibrated thermometer next to the tip of 

the Whirlpool thermometer. 

 

3.  Turn on the turbine assembly. 

 

4.  If the reading on the Whirlpool thermometer is not the same as 

that of the calibrated thermometer, turn the external adjustment 
screw (a) so that the Whirlpool thermometer reading is identical 
to the calibrated thermometer.

 

5.  Re-calibration shall be performed at least every 12 months or 

as needed. 

 

 

 

Summary of Contents for E-15-M

Page 1: ...H P 15 MU P 22 M Extremity E 15 M E 15 MH E 15 MU E 22 M E 22 MU E 27 M E 45 M Hi Boy H 60 M H 75 M H 90 M H 105 M Lo Boy L 75 M L 90 M L 105 M Sports S 85 M S 90 M S 110 M Slant Back SB 100 M P O Box...

Page 2: ...Manufacturing a division of Acorn Engineering and a member of Morris Group International Any reproduction dissemination or distribution of this document in whole or in part without prior written auth...

Page 3: ...e 7 Contraindications 7 User Skills 7 Device Description 8 Options and Accessories 9 Facility Requirements 11 Instructions 12 Assembly Instructions 12 Operating Instructions 12 Draining Instructions 1...

Page 4: ...o drop off the waste equipment for recycling contact your local waste collection authority General Warning or Caution Indicates personal injury or damage to the equipment is possible Electric Shock In...

Page 5: ...improper operation If such use is necessary this equipment and the other equipment should be observed to verify that they are operating normally Use of accessories and cables other than those specifi...

Page 6: ...cuit Interrupter GFCI medical grade receptacle No parts of the Whirlpool including the power supply cord are repairable or replaceable by service personnel No modification of the Whirlpool is allowed...

Page 7: ...n which hydrotherapy is contraindicated Active infection anywhere in the body Open wounds Loss of temperature sensation Hydrophobia Heat or cold intolerance Poor balance Kidney failure Safety Effectiv...

Page 8: ...water from the pump to a drain hose mounted at the top of the turbine by an optional drain pump located below the tub or by opening up the drain valve at the bottom of the tub Four casters all of whic...

Page 9: ...he color of the powder coat Powder Coat Paint All except Slant Back LOGO Logo selected by customer All except Slant Back The following accessories are available Description Used on Models Instructions...

Page 10: ...maintenance see the Disinfecting and Cleaning Instructions section TTS2 Tank Top Seat for 610 mm 24 inch wide tubs H 105 M S 90 M S 110 M L 75 M L 90 M L 105 M ARS Arm Rest Support All Hang support o...

Page 11: ...licensed electrician Environmental conditions within those specified in the Specification section of this manual NOTE The emissions characteristics of this equipment make it suitable for use in indus...

Page 12: ...rough hose if discharge nozzle 5 is lower than water level 4 Turn the pressure control valve handle 11 fully clockwise then fully counterclockwise to completely lubricate the plunger assembly 5 For Wh...

Page 13: ...p 12 to the bottom of the tub c Place discharge nozzle 5 into the sink d Turn the turbine power switch 13 on e When water stops discharging from the nozzle turn off the switch f Drain remaining water...

Page 14: ...at least 10 2 cm 4 inches of water 2 Place the stem tip of a calibrated thermometer next to the tip of the Whirlpool thermometer 3 Turn on the turbine assembly 4 If the reading on the Whirlpool thermo...

Page 15: ...ermometer and turbine shall be disinfected and cleaned after each patient Disinfecting and Cleaning the Tub and Thermometer 1 Drain the tub following the draining instructions above 2 Verify that the...

Page 16: ...ause Remedy Loss in pressure Foreign material caught in pump inlet holes 1 Turn off and unplug turbine assembly 2 Remove foreign materials from bottom of pump Turbine pump is not submerged fully in wa...

Page 17: ...A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 www whitehallmfg com 17 4 Replace the fuses with 6525 608 000 for the 125V turbine and 6525 612 000 for the 230V turbine or equivalent 5 Replace the...

Page 18: ...90 L 105 344 9 91 1 398 6 105 3 445 9 117 8 Water flow rate 0 56 8 L min 0 15 US gallons min Electrical Input voltage 115 230 V 230 V models are for export only Line Frequency 50 60 Hz Rated Current...

Page 19: ...527 U S A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 www whitehallmfg com 19 Transportation and Storage 0 60 C 32 140 F Relative humidity 5 95 non condensing Atmospheric Pressure 70 106 kPa Ther...

Page 20: ...such as mishandling or improper care or abnormal use The repair or replacement of the defective materials shall constitute the sole remedy of the Buyer and the sole remedy of Whitehall under this war...

Reviews: