11
Facility Requirements
The following are needed prior to using your Whirlpool:
Clean water source for filling tub
A sink or floor drain
Medical-grade GFCI receptacle, installed by licensed electrician
Environmental conditions within those specified in the Specification section of this
manual
NOTE: The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment
(for which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.