VeriPulse IOS User Manual
31
Begin your scan by starting on the arch of chosen restoration. If you have selected to scan the
antagonist, you may start by scanning either arch.
5.4.1 Scanning the Maxillary Arch
To make the scanning process more comfortable
for maxillary scanning, you may remove the tip
by pressing the button (1) located on the bottom
of the handpiece and rotating the scanner tip
180 degrees (2).
Recommended Maxillary Scan Path
When scanning the Maxillary arch, we recommend
you scan in the following order:
1. Occlusal – 2. Buccal – 3. Palatal
IMPORTANT:
Before use, the scanner’s removable tip must be sterilized
using an autoclave. See below: Section 7.2 Cleaning and Sterilization
of Tip.
1. Scan the Maxillary Occlusal area
Occlusal Surface
First scan the
OCCLUSAL surface
from molar to molar, ensuring
full occlusal surface is captured for all molars and premolars.
This initial path will drive the cross-arch accuracy of the scan, so always
Stay flat on the
teeth
. It may be useful to angle the scanner slightly when you come to the incisor and
canine teeth.