8
Function
Spec (DFS100)
Pulse Width
Range:
1.0 to 50.0 ms
Accuracy:
±
0.1 ms
Voltage
Range:
Up to 5000 V
Accuracy:
±
(1 % of r 2 V)
Note:
5000V is the V
p
Sample rate
100 kHz (10
μ
s sample)
Maximum
Average
Power
12 W, equivalent to 1 defib. pulses of 360 J every
30 seconds
DFS Battery
Level
Low power alert
2.3
ECG Signal Specifications
Table 3: ECG Signal Specifications
Function
Spec (DFS100)
Waves
Ventricular Fibrillation - Coarse
Ventricular Tachycardia (≥
230 BPM)
Normal Sinus
Rhythm (NSR)
Set at 80 BPM, Vp-p = 3 mV
Sample rate
5 kHz (200
μ
s sample)
Amplitude
0
–
5 mV
(Vp-p 5mV)