WHALETEQ
DFS100
User Manual
Revision 2021-08-03
Page 1: ...WHALETEQ DFS100 User Manual Revision 2021 08 03 ...
Page 2: ...rams as is without warranty of any kind either expressed or implied including but not limited to the implied warranties of merchantability or fitness for a particular purpose This document could contain technical inaccuracies or typographical errors Changes are periodically made to the information herein these changes will be incorporated in future revisions of this document WhaleTeq Co LTD is und...
Page 3: ...m 13 4 Getting Started 14 4 1 PC System Requirements 14 4 2 Software Installation 14 4 3 USB Driver Installation 15 4 4 Microsoft Net Framework Installation 16 4 5 PC Software and Firmware Auto Update 16 5 PC Software Operation 17 5 1 Software Function Introduction 17 6 Standalone Operation 19 6 1 One Time Test Convert 19 6 2 Continuous Arrhythmia Test 19 7 Calibration and Validation 20 8 Troubles...
Page 4: ...3 12 Revision History 22 13 Contact WhaleTeq 22 ...
Page 5: ...cations 8 Table 4 AED Discharge LED 9 Table 5 On Off and Pattern Selection Button 10 Table 6 DFS Battery LED 10 Table 7 DFS100 Test System 21 Table 8 Optional Accessories 21 Table 9 Optional Calibration Service and Warranty Extension 21 Table 10 DFS100 Standard Package Content 22 Table 11 User Manual Revision History 22 ...
Page 6: ...5 List of Figures Figure 1 Top Side of DFS100 9 Figure 2 Front and Rear Side of DFS100 11 Figure 3 Back Side of DFS100 12 Figure 4 Wiring Diagram 13 Figure 5 PC Operation Interface 17 ...
Page 7: ... DFS100 is the ideal tool to conduct such a regular inspection 1 2 Applications WhaleTeq DFS100 is a compact field tester functioning as an ECG signal simulator and defibrillation energy guarantee equipment for AED regular maintenance For ECG patterns simulation DFS100 simulates NSR Normal Sinus Rhythm as well as other arrhythmias that may need AED defibrillation Arrhythmia convert combination ECG...
Page 8: ... 9V Battery 1 no charger Mechanical Housing ABS Plastic Size L x W x H 17 0 x 8 55 x 4 0 cm Weight 330 g Safety Standards CE IEC EN61010 1 2010 A1 2016 Pollution degree 2 61010 2 030 2017 Electromagnetic Compatibility Standards EMC FCC EMC P15B 2 2 Energy Measurement Specifications Table 2 Energy Measurement Specifications Function Spec DFS100 Maximum Energy Up to 600 J Accuracy 1 of reading 1 J L...
Page 9: ...e Maximum Average Power 12 W equivalent to 1 defib pulses of 360 J every 30 seconds DFS Battery Level Low power alert 2 3 ECG Signal Specifications Table 3 ECG Signal Specifications Function Spec DFS100 Waves Ventricular Fibrillation Coarse Ventricular Tachycardia 230 BPM Normal Sinus Rhythm NSR Set at 80 BPM Vp p 3 mV Sample rate 5 kHz 200 μs sample Amplitude 0 5 mV Vp p 5mV ...
Page 10: ...his side up when in use B AED Discharge LED Table 4 AED Discharge LED Scenario LED behavior No Discharge No light Discharge Detected Pass If the discharge value equals to or is greater than the joule number set in 5 1 C the green light continues for 30 seconds during which Pattern Selection Button D has no function A E A B C F D ...
Page 11: ...s the button for about 1 second to turn on DFS100 Pattern Selection When DFS Battery LED F indicates DFS100 is on the button function becomes pattern selection Off Long press the button for 3 seconds to turn off DFS100 Alternatively after DFS100 idles for 3 minutes it ll automatically be turned off The length of idle time can be customized via PC software described in 5 1 D E Mode Switch a Mode 1 ...
Page 12: ...ED pad labeled Sternum via DFS accessory AED Interface Cable I Caution DFS100 is recommended to be used for analyzing defibrillation shock at the interval of at least 60 seconds apart J Caution the possibility of electric shock From the time AED starts charging for defibrillation until right after energy delivery User is recommended to stay clear from DFS100 K USB Port Connect to the computer s US...
Page 13: ...ide of DFS100 L Label Contain information including model serial number manufacturer and power supply M Battery Compartment Insert a 9V battery for power supply Both disposable batteries and rechargeable batteries can be used L M ...
Page 14: ...13 3 4 Wiring Diagram Figure 4 Wiring Diagram Note To plug in properly please make sure the yellow lines on the black and red banana males that connect to DFS100 totally disappear in the jacks ...
Page 15: ...Double click on WhaleTeqDFS100 exe to execute DFS100 PC software Note If DFS software cannot be executed properly or this is the first time using WhaleTeq s product please refer to the two sections below 4 3 and 4 4 to confirm that the USB driver and Microsoft Net Framework 4 0 are both installed 1 Relative to normal PC processing there is no special use of PC speed However there has been noted a ...
Page 16: ...ch have built in USB functions As mchpcdc inf provided by Microchip does not contain a digital signature please disable the driver signature enforcement in Windows 8 and Windows 8 1 Please click here to watch the tutorial video When the USB module is connected for the first time select Manual Installation and point to the folder containing the above file Then continue to follow the instructions to...
Page 17: ...q s software is developed based on Microsoft Net Framework 4 0 If DFS software fails to execute properly please check whether Microsoft Net Framework 4 0 or a higher version was installed in the operating system If the PC hasn t installed Microsoft Net Framework 4 0 or a higher version please download it from Microsoft s website Please click here to watch the tutorial video from 2 03 4 5 PC Softwa...
Page 18: ...es so upon 30 seconds after receiving defibrillation Note If User doesn t enable the Auto Return to Initial ECG Pattern function User will need to manually switch the pattern from NSR to VT or VF if needed for the next inspection C Joule Threshold Depending on different models of DUT AED the delivered defibrillation is different DFS100 allows User to self define the passing criteria of the defibri...
Page 19: ...Check for Updates Click to see if the latest version of firmware and or software has been released Please note that the USB cable used for firmware upgrade must be file transferrable F About Check the software version and the device function introductions ...
Page 20: ...t fails the LED lights red At the same time DFS100 simulates the patient having been resuscitated by changing the ECG pattern to NSR 6 2 Continuous Arrhythmia Test In this mode DFS100 continuously sends user selected arrhythmia to the AED Ensure the AED and DFS100 are connected properly via the interface cable Turn on both DFS100 and the AED and follow the AED instructions The default setting of D...
Page 21: ...t convenience 9 Cautions 1 Follow AED instructions during testing with DFS100 2 Follow the color code and or labeling when connecting DFS100 and the AED via the AED interface cable 3 When DFS100 is in use the face containing LEDs and buttons must face up to allow ventilation 4 To allow ventilation DFS100 should be used to analyze defibrillation energy at the interval of at least 60 seconds 5 USB c...
Page 22: ...21 9V alkaline battery K22 0500101 Open ended AED cable Banana Male 0 5 meter Black K22 0500102 Open ended AED cable Banana Male 0 5 meter Red N61 0210033 Compact splicing connector Customization of AED interface cable adapter is available Please contact WhaleTeq service whaleteq com Table 8 Optional Accessories Order No Description K27 1800304 USB transfer cable A to C M to M 1 8 meter G34 060010...
Page 23: ...Banana Male 0 5 meter Red 1 Compact splicing connector 2 DFS100 Optional Accessories USB transfer cable A to C M to M DFS carrying bag 12 Revision History Table 11 User Manual Revision History Version Revision Description Issue Date 2021 08 03 First Version 2021 08 13 13 Contact WhaleTeq WHALETEQ Co LTD service whaleteq com O 886 2 2517 6255 8F No 125 Songjiang Rd Zhongshan Dist Taipei City 104474...