CMRR 3.0E | User Manual
18
CMRR 3.0E connection status
CMRR 3.0E test parameter display
CMRR 3.0E test parameter adjustment
Send setting parameters to CMRR 3.0E
Battery capacity of DC offset
Launch CMRR 3.0E Assistant software
Figure 6-6, CMRR 3.0E PC software UI
6.2.2
CMRR 3.0E Assistant Software
Refer to figure 6-7 for software interface of CMRR 3.0E Assistant Software. This is mainly used to support
test sequences required by IEC80601-2-26:2019. CMRR 3.0E Assistant Software can effectively simplify
the complication of switching different test parameters.
Once select IEC-2-26 standard, follow the software guide to go through the setting of test parameters.
Also, the test can be conducted manually or automatically.
Here we take IEC-2-26 as an example to explain the test sequences.
6.2.2.1
Step 1. Preparation
First, after clicking [IEC-2-26], the test step automatically starts at "Step 1. Preparation". It also explains
required settings, as shown in the three green-shaded descriptions in figure 6-7:
1.
No EEG patient cable is attached
2.
The line frequency notch filter (if provided) of EEG is turned off
3.
Set the ECG GAIN to 0.1 mm/uV
Make sure EEG to follows these three settings, and click [Next] to continue to next step.