background image

Operation Manual

for Westover Model RHB-18

Ô

1)

Open daylight plate, and place

2-3 drops of distilled water on
the main prism. Close the
daylight plate so the water
spreads across the entire
surface of the prism without air
bubbles or dry spots. Allow the
sample to rest on the prism for
approximately 30 seconds
before going to step #2. (This
allows the sample to adjust to
the ambient temperature of
the refractometer)

2)

Hold daylight plate in the

direction of a light source and
look into the eyepiece. You
will see a circular field with
graduations down the center
(you may have to focus the
eyepiece to see the
graduations clearly). The
upper portion of the field
should be blue, while the
lower portion should be white.

Poor

Poor Good

18
16
.
.
.
4
2
0

18
16
.
.
.
4
2
0

3)

While looking into the eyepiece,

turn the Calibration Screw until
the boundary between the
upper blue field and the lower
white field meet exactly on the
0.0°. Once complete, the
instrument is calibrated for
your current ambient room
temperature. When working
temperature of the room or
environment (not the sample)
changes by more than 5°F,
we recommend recalibrating
to maintain accuracy and
reproducibility.

As seen when looking

into the instrument

Upper Blue Field

Lower White Field

Boundary Line

adjust to here

Calibration Procedure

Parts Diagram

Daylight Plate

Calibration

Screw

Focus

Adjustment

Main Prism

Assembly

Rubber

Grip

Eyepiece

Hand Held Brix Refractometer

Range:

0-18°

Minimum Division:

0.2°

Accuracy:

0.1°

Dimensions:

27 x 40 160mm

Weight:

176 grams

Reviews: