background image

• 

1.  Place the calibration tool into 

the holes of the calibration 

adjustment lock (under the  

push button). 

2.  Turn the adjustment lock 

anticlockwise to decrease and 

clockwise to increase the volume. 

3.  Repeat checking calibration until 

the pipetting results are correct.

12

PRODUCT MANUAL

1800 358 101

[email protected]

www.westlab.com.au

+64 9 553 3677

[email protected]

www.westlab.co.nz

646 568 5391

[email protected]

www.westlab.com

1 877 822 1455

[email protected]

www.westlab.com

Australia

New Zealand

United States

Canada

Autoclaving

The pipettes are fully autoclavable by steam at 121°C, 1.05 bar for 15 min.  

Autoclave the complete pipette without any disassembling. 

After autoclaving under the mentioned conditions, allow the pipette to cool and dry  

for 6 hours before use. It is not required to re-calibrate the pipette after autoclaving. 

If the pipette is autoclaved frequently, the piston and springs should be greased with 

the lubricant supplied with each pipette to maintain smooth movement.

Calibration 

Performance testing should take place in a draught-free, stable environment with a 

relative humidity above 50% and constant (±0,5 °C) temperature between 15°C and 30°C. 

The weighing vessel, pipette, tips, and distilled water need to have been in the room for a 

sufficient time, at least 2 hours, to reach equilibrium with the room conditions.

Use an analytical balance with a readability of 0.01 mgs. 10 measurements for each test 

volume shall be carried out. For variable volume pipettes, at least three volumes shall be 

tested (nominal volume, 50% of nominal volume, and 10% of nominal volume).

1.  Select the test volume. Do not change the setting during the test cycle of  

10 measurements.

2.  Fit the selected tips to the pipette cones.
3.  Fill the tips with distilled water and condition the pipette before testing by aspirating 

and dispensing the distilled water in the tips five times. Afterwards discard the tips. 

4.  Attach new tips on the pipette cones and pre-wet the tips once.
5.  Aspirate the distilled water and pipette it into the weighing vessel  

(use forward pipetting technique).

6.  Weigh the pipetted quantity with an analytical balance and record the weight. 

Repeat the test cycle until 10 measurements have been recorded.

7.  Convert the recorded masses to volumes, calculate the mean volume,  

systematic error and random error as described in EN ISO 8655-6:2002, part 8.

8.  Compare the systematic error (inaccuracy) and the random error (imprecision) 

with the values in the specification table. 

The calibration of the pipette must be set even if only one of the results falls outside 

the permitted range.

For checking the calibration of a multichannel pipette, each channel should be 

considered as an individual pipette.

Recalibration 

Summary of Contents for 665-259

Page 1: ...1 PRODUCT MANUAL Precision Multichannel Pipettes Codes 665 259 665 260 665 261 665 262 665 263 665 264...

Page 2: ...overall quality of the pipette Furthermore the pipettes are manufactured from a heat resistant material allowing them to be safely autoclaved multiple times without a loss of quality With a fully auto...

Page 3: ...annel pipette Silicone grease Calibration assembling tool Tool to remove colour coding drop Bag of drops for colour coding Bag of sample tips Bag of O rings Instruction manual Calibration report 2 Tec...

Page 4: ...es that are not supplied with the unit Is operated with accessories that are not supplied or recommended by the manufacturer Is operated improperly or contrary to the manufacturer s specifications 4 M...

Page 5: ...ssembly parts Number Name 1 Push button 2 Top closure 3 Spring stopper 4 Primary spring support 5 Primary spring 6 Main body 7 Counter Number Name 8 Washer 9 Shaft 10 Bottom closure snap fitting 11 Ti...

Page 6: ...ab com www westlab co nz 646 568 5391 sales us westlab com www westlab com 1 877 822 1455 sales ca westlab com www westlab com Australia New Zealand United States Canada Bottom assembly parts Number N...

Page 7: ...ia New Zealand United States Canada Calibration disassembling tool The calibration disassembling tool is supplied for maintenance and calibration of the pipette Pipette identification The pipettes are...

Page 8: ...tool between the tip ejector and the tip ejector collar to release the locking mechanism 3 Carefully release the tip ejector and remove the ejector collar 4 Place the wrench end of the opening tool o...

Page 9: ...spirating and dip the tips only 2 3mm into the liquid Pre rinse the tips five times before aspiration by filling and emptying the tips Always control the push button movements with the thumb for consi...

Page 10: ...ip display window 2 The desired volume is set by turning the push button clockwise or anticlockwise ATTENTION Using excessive force to turn the push button may jam the mechanism and damage the pipette...

Page 11: ...will empty the tip completely and ensure accurate pipetting 5 Release the push button to the ready position Reverse pipetting Reverse pipetting is recommended for viscous solutions with a tendency to...

Page 12: ...m for a sufficient time at least 2 hours to reach equilibrium with the room conditions Use an analytical balance with a readability of 0 01 mgs 10 measurements for each test volume shall be carried ou...

Page 13: ...etween the tip and the tip cone Clean the tip cone Insufficient amount of grease on the piston Clean and grease the piston Inaccuracies Incorrect operation Follow instructions carefully Calibration al...

Page 14: ...lab com au 24 hours 7 days a week convenience 14 PRODUCT MANUAL 1800 358 101 sales au westlab com www westlab com au 64 9 553 3677 sales nz westlab com www westlab co nz 646 568 5391 sales us westlab...

Page 15: ...15 PRODUCT MANUAL This page has been left blank intentionally...

Page 16: ...Westlab in every discovery...

Reviews: