1
PRODUCT MANUAL
Product Manual
Westlab Digital Stirred Water Baths
Codes 663-898, 663-899, 663-900, 663-903, 663-904
Page 1: ...1 PRODUCT MANUAL Product Manual Westlab Digital Stirred Water Baths Codes 663 898 663 899 663 900 663 903 663 904...
Page 2: ...rrer is quiet in use and provides excellent temperature stability Indication of heating over temperature alarm and timer running When a low liquid level is detected a FILL message automatically appear...
Page 3: ...e Ambient 5 C to 99 C Temperature Alarm Digital LED Display Under Over Temperature Alarm is Automatically Set 4 C Above 4 Below Set Point Temperature Control PID Digital LED Display Resolution of 0 1...
Page 4: ...he Maintenance section to ensure the proposed method will not damage the unit 8 Connect only to a power supply with the corresponding voltage to that specified on the rating label positioned on the re...
Page 5: ...within your laboratory ATTENTION Disconnect the Bath from the Power Supply Prior to Cleaning Basic Cleaning Procedure The stainless steel crevice free tank should provide years of valuable service an...
Page 6: ...alse base into the bath with the legs down Fit the power lead into the socket at the rear 4 2 Operation Switch the unit on using the mains switch located on the rear The switch will illuminate and the...
Page 7: ...imed period has finished Press the RUN key to deactivate the beep and clear the END message Power Interrupted During Timer Mode If the power is interrupted during the timer mode the display shows P OF...