TROUBLESHOOTING
If any abnormality arises during use, please check the following points.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn on
when the
START/STOP
button is pushed.
Batteries have run
down.
Replace them with four new
“AAA” alkaline batteries.
Battery polarities (+/-)
have been positioned
incorrectly.
Re-insert the batteries in the
correct positions. See page 11.
EE measurement error
symbol shown on
display or the blood
pressure value is
displayed excessively
low (or high).
The UltraSoft™ upper
arm cuff has been
positioned on the arm
incorrectly.
Re-wrap the cuff properly so
that it is positioned correctly.
Take new measurement.
Did you talk or move
during measurement?
Keep arm steady during
measurement. Measure
again. Refer to “Measurement
Procedure” instructions, see
page 17.
Shaking of the arm with
the cuff on.
E1 error symbol shown
on Display.
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Check cuff connection.
See page 15. Take new
measurement.
E2 error symbol shown
on Display.
Inflation pressure
exceeding 300 mmHg.
Switch the unit off and then
turn back on. Take new
measurement.
E3 error symbol shown
on display.
Error determining
measurement data.
Re-wrap the cuff properly so that
it is positioned correctly. Take
new measurement.
EP error symbol shown
on display.
System Error.
Remove batteries. After 1
minute, re-insert the batteries.
Take new measurement.
Power Source:
4 1.5V DC “AAA” alkaline batteries
or 6V DC AC adapter
Measurement Method: Oscillometric
Measurement Range:
Pressure: 40~280 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
100 memories for each User (200 total)
Auto-shut-off:
1 minute after last button operation
Operation
Environment:
Temperature: 41°F~104°F (5°C ~40°C)
Humidity: 15% ~ 93% RH
Storage/Transportation
Environment:
Temperature: -13°F~158°F (-25°C~70°C)
Humidity: Less Than 93% RH
Weight:
0.58 lbs (265 g) (without batteries)
Arm Circumference:
Dimensions:
4.49”(L) x 5.63”(W) x 1.79”(H)
114mm(L) x 143mm(W) x 45.5mm(H)
Accessories:
6V DC AC adapter, 1 arm cuff with tube,
intruction manual, 4 “AAA” alkaline batteries
Internally Powered equipment
Type BF Equipment
IP22-Degrees of protection provided by
enclosures
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or
nitrous oxide.
Continuous operation with short-time loading.
Follow instructions for use.
SPECIFICATIONS
Note: These specifications are subject to change without notice.
UltraSoft
TM
Upper Arm Cuff size: 9”–17” (23-43 cm)
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair the
unit by yourself. Contact information for HoMedics Consumer Relations
Department can be found on the warranty page.