- 20 -
Press
to exit.
Probe type
C1-11/50R/3.5MHz
C1-12/20R/5.0MHz
angle
70º visual and adjustable
96º visual and adjustable
4.10 Focus adjustment
Focus number
Press 7 or
8 button in real-time B mode or BB mode state to light “Focus” on the top right of screen,
and then press “
+/9
”button repeatedly to switch the focal point number to one or two. Meanwhile, the
current focal point number is displayed on the right of screen, such as “focus: 2 D3”, “D3” means that the
focal distance is 3 cm.
Press
to exit.
Attention:
When the focus is set to 2, the current image frame frequency would be reduced.
Focal position
Press 7 or
8 button in real-time B mode or BB mode state to light “focus” on the top right of screen,
and then press ▲5 and
▼6
button to move the focus up and down, and change the focal position.
Press “
”→“
B/0
”in turn to exit.
Tip
:
The focusing of image near the focus indicator
is better, and you can change the focus
number and focal position, so that the focus will point to the interest areas.
4.11 probe work frequency
Press 7 or
8 button in real-time state to light “frequency” on the top right of screen, and then press
▲5 and
▼6
button to adjust frequency. The current frequency will be displayed on the top right of screen,
such as “frequency: 3.5 MHz”. Frequency range of every kinds of probe is listed as following:
LV2-4/6.5MHz rectal probe —— 5.0MHz, 5.5MHz, 6.5MHz, 7.5MHz
C1-11/50R/3.5MHz convex array probe —— 2.0MHz, 2.5MHz, 3.5MHz, 5.0MHz
C1-12/20R/5.0MHz convex probe—— 4.0MHz , 4.5MHz, 5.0MHz, 5.5MHz
L1-5/7.5MHz high frequency linear-array probe —— 6.0MHz , 6.5MHz, 7.5MHz, 8.5MHz
35LV150CP Backfat probe
——
2.5MHz
、
3.5MHz
、
4.5MHz
、
5.5MHz