SETUP

1.

Install the supplied Oral/Axillary probe or optional Rectal probe by inserting
the latching plug on the end of the coiled cord into the plug receptacle on the
unit.  The plug snaps into place and cannot be removed without  depressing
the latch.    (Figure 1).

Note:  The Oral probe is also used to
take Axillary measurements.

2.

Observe the LCD to see that the probe
type OrL = Oral, ALy = Axillary, or rEC
= Rectal is displayed.

3.

Insert probe shaft completely into probe
storage channel.

4.

Remove the probe from the storage
channel.

5.

Observe the LCD to see that all display
segments appear followed by the display
of the probe type OrL = Oral , ALy =
Axillary, or rEC = Rectal and tempera-
ture scale (°C or °F).  Replace probe in
storage channel.

6.

Insert a box of disposable probe covers (part number 05031) into the
probe cover storage well.

CHANGING PROBES

• Depress the latch and unplug the probe connector from the thermometer.

• Remove the probe from the storage channel.

• Repeat steps 1 through 6 in the SETUP section.

FEATURES

Normal Mode:  The thermometer takes oral temperatures in about 4
seconds and rectal temperatures in about 15 seconds.  Axillary tempera-
tures taken using the Normal operating mode will display temperatures in
about 10 seconds for  patients under 4 years.  The unit will always operate
in Normal mode unless the Mode button is depressed for more than 2
seconds after the initial temperature is displayed, which changes it to
Monitor Mode.  Replacing the probe in the storage channel automatically
resets the unit to Normal mode.

8

High Temperature:  This icon is displayed if the patient’s temperature
is greater than 108.0°F (42.2°C).

Low Temperature:  This icon is displayed if the patient’s temperature
is less than 84.0°F (28.9°C).

High Ambient Temperature:  This icon is displayed if the ambient

temperature is greater than 104°F (40°C).

Low Ambient Temperature:  This icon is displayed if the ambient

temperature is less than 60.8°F (16.0°C).

Low Battery:  This icon flashes when batteries are low and will
remain on when batteries are dead.

Security Mode: Return the thermometer to the wall holder in order
to resume normal operation.

Display Test:  All segments of the thermometer display will appear
briefly when the probe is removed from the storage channel to ensure
proper function.

Summary of Contents for SureTemp 678

Page 1: ...tely returned securely packaged and postage prepaid to Welch Allyn Loss or damage in shipment shall be at purchaser s risk Welch Allyn will not be responsible for loss associated with the use of any Welch Allyn productthat 1 hashadtheserialnumberdefaced 2 hasbeenrepairedbyanyoneotherthan an authorized Welch Allyn Service Representative 3 has been altered or 4 has been used in a manner other than i...

Page 2: ... including photocopy without prior consent in writing from Welch Allyn Printed in the U S A U S patent 5 632 555 Additional U S and foreign patents pending The CE mark on this product indicates it has been tested to and conforms with the provisions noted within the 93 42 EEC Medical Device Directive European Regulatory Manager Medical Division Welch Allyn Ltd Navan Co Meath Republic of Ireland 15 ...

Page 3: ...ndentations located on the sides of the rear housing This will release two snaps 2 Remove the battery door 3 Remove batteries by lifting the battery strap IMPORTANT After removing the old batteries press and hold the MODE button for 5 seconds before inserting the new batteries 4 Replace batteries with 3 AA Alkaline only Orient per markings on the inside of the battery well 5 Replace the battery do...

Page 4: ...seoflubricantisoptional WARNING Incorrect insertion can cause bowel perforation Tilt the probe so the tip is touching tissue Keep hand separating buttocks in place Hold the probe in place during the entire temperature measurement process 4 When the final temperature is reached a tone will sound and the temperature will be displayed Rectal temperature mea surements will take about 15 seconds 5 Afte...

Page 5: ...n button are to be used for taking oral and axillary temperatures only Rectal probes red ejection button are to be used for taking rectal temperatures only The use of the wrong probe will produce temperature errors The thermometer case is not waterproof do not drip fluids onto it Do not use this instrument for any purpose other than that specified in this booklet Doing so will invalidate the instr...

Page 6: ... the Axillary Mode with an oral probe 1 Withdraw the oral probe from its storage channel and observe a quick thermometer self test on the display 2 With OrL probe type displayed depress and hold the Mode button for two seconds ALy will now appear in the display indicatingthat the thermometer is ready to take an axil lary temperature mea surement Unless the thermometer is in the axillary mode inacc...

Page 7: ...ll holder Note Replacingtheprobeintotheprobe holder will place the thermometer back into its normal operating state Once in the Biotech Mode there are several functions that can be accessed These function are described completely in the Model 678 Technical Manual The following section describes the security functions only 10 3 During the temperature measurement cycle walking segments will appear o...

Page 8: ... On Security Feature With this function set to ON the thermometer immediately goes into security alarm when removed from the security wall holder The options here are on and off In the OFF setting the unit will not alarm when removed from the wall holder Once the security functions have been set to the desired state replace the probe back into the probe holder to exit the Biotech mode 9 OPERATION ...

Page 9: ...al temperatures in about 4 seconds and rectal temperatures in about 15 seconds Axillary tempera tures taken using the Normal operating mode will display temperatures in about 10 seconds for patients under 4 years The unit will always operate in Normal mode unless the Mode button is depressed for more than 2 seconds after the initial temperature is displayed which changes it to Monitor Mode Replaci...

Page 10: ... probe storage channel ModeTable 7 DISPLAYINDICATORS Instrument Malfunction This icon is displayed if the unit is not functioning properly Contact the Welch Allyn Customer Service Department Broken Probe If displayed probe may be damaged Remove and reinstall probe If error persists then install a new probe Probe Position Probe has broken contact with the tissue Once proper contact has been achieve...

Reviews: