v
v
Power Supply
The mains connection is on the rear of the unit.
The power supply voltage is set by the factory for100-115V(nom. 110V) or 220-240V
(nom. 230V) working.
The setting is indicated by the indented metal strip on the fuse
panel. Contact your dealer if the voltage needs to be changed.
The mains indicator lamp on the keyboard is always lit when the unit is connected to the mains
supply. The unit can either be operated from the mains supply or from the built-in recharge-
able battery. The power source is indicted on the top line of the LCD.
Changing a Mains Fuse
If it is necessary to change a fuse, always replace with the correct rating i.e 2x200mAT
for 230V, or 2x315mAT for 110V .
To change a fuse press the two retaining lugs on side of the fuse panel (situated below the mains
connector on the back panel. Remove the fuse panel and replace the fuse(s). Click back the fuse
panel.
Power Supply
The mains connection is on the rear of the unit.
The power supply voltage is set by the factory for100-115V(nom. 110V) or 220-240V
(nom. 230V) working.
The setting is indicated by the indented metal strip on the fuse
panel. Contact your dealer if the voltage needs to be changed.
The mains indicator lamp on the keyboard is always lit when the unit is connected to the mains
supply. The unit can either be operated from the mains supply or from the built-in recharge-
able battery. The power source is indicted on the top line of the LCD.
Changing a Mains Fuse
If it is necessary to change a fuse, always replace with the correct rating i.e 2x200mAT
for 230V, or 2x315mAT for 110V .
To change a fuse press the two retaining lugs on side of the fuse panel (situated below the mains
connector on the back panel. Remove the fuse panel and replace the fuse(s). Click back the fuse
panel.
Summary of Contents for Schiller AT-2plus
Page 1: ...i i AT 2plus Spiro Option AT 2plus Spiro Option Schiller Menu Manuals...
Page 9: ...ix ix...
Page 10: ...x x...