Service Manual
Repair Procedures
57
7.
Insert the ECG sensor cable into the J1 connector and lock the latching piece.
8.
Install the screws and rectangular washers to secure the main board to the front
case. Torque the screws to 3 in. lbs.
Closing the Case
1.
Insert the rubber gasket (part of the fastener kit) into the groove in the front case.
Begin inserting the gasket in the end locking section, located on the lower edge of the
case (as shown below).
Caution
Do not to bend or tilt any of the neon bulbs. They must remain vertical.
Smooth curve in flex cable
Correct
Incorrect
Sharp bend in flex cable
ECG sensor flex cable at J1
Neon bulbs
Screws and rectangular washers