Service Manual
Repair Procedures
49
Disassembly - Model 402
This procedure covers the Model 402 monitor with ECG only.
Opening the Case
1.
Remove the battery pack.
2.
Remove the rubber feet from the rear panel.
3.
Remove the six case screws.
4.
Separate the case.
Note
If you are replacing only the rear case, proceed to
Case screws