background image

990154 Rev. A

5

 

8VLQJWKH6\VWHP

Rea dy

1.  Turn the unit on. All of the lights on the Control

Panel will flash and the yellow Standby indicator will
turn on. When the lights stop flashing, the unit is in
Standby mode. The values shown will be the
settings from the last use.

Note: Opening the lid of the unit will depressurize

the system and power will turn off.

2. Press the 

▲▼

 keys to adjust the Particle Energy to

provide optimum powder flow. This adjustment can
be made in either Ready or Standby mode. A low
Particle Energy setting (50-70) is suggested when
first using the instrument.

3.  Set the Beam Intensity (1-10) by pressing the Beam

Intensity 

▲▼

 keys. The suggested beginning setting

for Beam Intensity is 5.

4. Press A i r  to select the Ready mode.

5.  Lift the handpiece from the cradle. If the handpiece

was out of the cradle during power up, it must be
replaced and then removed prior to pressing the
footswitch. Failure to do so will result in an error.

WARNING: Inspect the condition of the handpiece

tip prior to each use.

WARNING: Direct the initial powder sample into the

high volume evacuation system prior to each
patient use or when changing system settings.

6.  Press the footswitch to start the powder flow through

the nozzle.

Note: If the handpiece is out of the cradle during power

up, it will need to be replaced and then removed
prior to pressing the footswitch.

Unit Shutdown

1. Press A i r  to select Standby mode.

2.  Set the Power Switch to OFF.

Note: If the handpiece is left out of the cradle and the

unit is not used for 10 minutes, the unit will revert
to Standby mode. When in Standby or when the
handpiece is placed in the cradle, the footswitch
is disabled. The unit will go to Standby when the
handpiece is placed in the cradle.

 

0DFK6SHFLDO,QVWUXFWLRQV

The Mach 8 is a multi-operatory version of the Mach 7.
It is configured to use the main unit display as one
operatory, with the display control module (included)
operating from a remote site up to 16 feet away.

An optional second remote can be connected to
provide dual remote functions with separations up to 32
feet. Contact your authorized Welch Allyn / Kreativ
representative for these optional accessories.

I / O

Rea dy

The Mach 8 display panel and controls are the same as
the Mach 7, except the Mach 8 has two handpiece
hose connections. Before using the unit, verify that
each hose works with its corresponding control panel
and foot switch.

Summary of Contents for Kreativ

Page 1: ...torage and Shipment 9 10 Specifications 10 11 Recycling 10 CAUTION This device must only be purchased by or sold to a licensed dentist Clinical use of the Mach 7 8 Air Abrasion system for procedures not included or suggested in this guide is not advised The Mach 7 8 Air Abrasion system should only be used by dental professionals who have practiced on extracted teeth and are thoroughly familiar wit...

Page 2: ...d Do not attempt to service the unit beyond those items described on page 6 Maintenance Call an authorized Welch Allyn Kreativ service representative for instructions Practice on extracted teeth prior to using the Mach 7 8 on patients Begin with the 27 5 micron aluminum oxide powder Use only short bursts carefully monitoring your progress Most users prefer the 27 5 micron powder at a Particle Ener...

Page 3: ...and other materials to prepare teeth for restoration and as an adjunct to a high speed dental handpiece and bur The system delivers a controlled mixture of abrasive particles in a dry air stream through the handpiece nozzle This manual should be thoroughly reviewed before installing and or using the equipment Please call Welch Allyn Kreativ or their representative if you need additional informatio...

Page 4: ...ded with your unit refer to your remote control User s Guide for connection Note A remote plugged into a Mach 7 will take control from the front panel A remote plugged into a Mach 8 will function as a second operatory Note A special footswitch Y adapter cable is required for the Mach 8 see page 5 CAUTION Do not exceed 135 PSI at input Applying pressure in excess of 135 PSI will cause the relief va...

Page 5: ...prior to each patient use or when changing system settings 6 Press the footswitch to start the powder flow through the nozzle Note If the handpiece is out of the cradle during power up it will need to be replaced and then removed prior to pressing the footswitch Unit Shutdown 1 Press A i r to select Standby mode 2 Set the Power Switch to OFF Note If the handpiece is left out of the cradle and the ...

Page 6: ... case and control areas with a soft cloth slightly moistened with a mild detergent or isopropyl alcohol Do not use any type of abrasive cleaners scouring powder or solvents on the case Case inside Excessive powder inside the equipment may be removed with a vacuum or a cloth slightly moistened with water or isopropyl alcohol Never use a moistened cloth on the Printed Circuit boards Powder Chamber C...

Page 7: ...n the pinch tube assembly with the grooves and securely tighten the thumb screws Monthly Requirements Internal Tubing 1 Inspect internal tubing for wear and air leakage 2 Listen for audible leaking sounds and look for powder accumulation internal to the unit 3 Soapy water and a small paintbrush are ideal for finding small air leaks 4 Clean up excess powder to help isolate the leak area 5 If a leak...

Page 8: ...e Ruptured handpiece hose Call local service rep Air flow from nozzle when footswitch is not depressed Loss of air pressure due to a misadjusted pinch valve regulator Put system in Standby and call local service rep Faulty footswitch Put system in Standby and call local service rep Defective pinch valve Put system in Standby and call local service rep Pinch valve block is not tight Put system in S...

Page 9: ...ir abrasive tubing hand pieces and handpiece tips are not warranted against normal wear and tear or customer damage This limited warranty applies only to the original purchaser user of the equipment The foregoing limited warranty is exclusive and in lieu of all other warranties whether written oral or implied and shall be purchaser s sole remedy and Welch Allyn Kreativ s sole liability under contr...

Page 10: ...Allyn Kreativ 5HF FOLQJ All recycling must be in compliance with relevant national laws Government authorized recycling centers must be used There are no hazardous materials in this product Device All covers and housings should be recycled steel or aluminum Plastic parts should be removed from the metal pads before recycling e g power inlet footswitch connector main switch etc All printed circuit ...