990154 Rev. A

10

 

6SHFLILFDWLRQV

Power

100-240 V~, 50 / 60 Hz,
0.75A - 0.35A using external DC
power supply delivered with unit.

Air

80 psi to 130 psi (550 kPa to 896
kPa) with at least 1.2 cfm (34
l/min) recommended

Size

10” x 15” x 10” (25cm x 38cm x
25cm)

Weight

12 lbs. (5.5 kg)

Environment

Operational

Temp: 50

°

F (+10

°

C) to 104

°

F

(+40

°

C)

Relative humidity: 30% to 75%

Pressure: 10.15 psi (70 kPa) to
15.3 psi (106 kPa)

Transport and

Storage

Temp: -40

°

F (-40

°

C) to 158

°

F

(+70

°

C)

Relative humidity: 10% to 90%

Pressure: 7.25 psi (50 kPa) to
15.3 psi (106 kPa)

Equipment

Classification

Class I (requires an electrical
earth ground) and ordinary
protection against ingress of
water
Applied Part: Type B
Mode of Operation:

Continuous Operation

Particle

Energy

Accuracy

The Particle Energy (PSI) display
is an internal pressure
measurement and should not be
interpreted as the precise
pressure at the handpiece nozzle.

Electrical

Fuses

There are no customer
serviceable fuses.

Options

A wide selection of abrasive
powders and nozzle tip angles
and sizes are available from
Welch Allyn / Kreativ.

 

5HF\FOLQJ

All recycling must be in compliance with relevant
national laws. Government authorized recycling centers
must be used. There are no hazardous materials in this
product.

Device

 

All covers and housings should be recycled
(steel or aluminum).

 

Plastic parts should be removed from the metal
pads before recycling (e.g., power inlet, footswitch
connector, main switch, etc.)

 

All printed circuit boards should be recycled as
electronic devices.

Accessories

 

All spare parts should be recycled as listed above.

 

Parts that have been in contact with the patient
should be cleaned and sterilized.

Packaging

The carton, all other plastics, and inserts should be
recycled (cardboard, foam, polyethylene).

Environmentally Relevant Materials

Components

Materials

Enclosure, all brackets

Steel, Aluminum, ABS

Printed Circuit Boards

Epoxy, FR4

SAVE THE CARTON!  The Mach 7/8 and all
accessories are shipped in a single container. The
shipping container and packing materials should be
retained in the event that the instrument has to be
returned for service or repair.

Manufactured by
Welch Allyn / Kreativ, Inc. USA
7420 Carroll Road
San Diego, CA  92121

European Regulatory Manager

Welch Allyn LTD.

Navan, Co. Meath

Republic of Ireland

Ph: 353-46-79060

Summary of Contents for Kreativ

Page 1: ...torage and Shipment 9 10 Specifications 10 11 Recycling 10 CAUTION This device must only be purchased by or sold to a licensed dentist Clinical use of the Mach 7 8 Air Abrasion system for procedures not included or suggested in this guide is not advised The Mach 7 8 Air Abrasion system should only be used by dental professionals who have practiced on extracted teeth and are thoroughly familiar wit...

Page 2: ...d Do not attempt to service the unit beyond those items described on page 6 Maintenance Call an authorized Welch Allyn Kreativ service representative for instructions Practice on extracted teeth prior to using the Mach 7 8 on patients Begin with the 27 5 micron aluminum oxide powder Use only short bursts carefully monitoring your progress Most users prefer the 27 5 micron powder at a Particle Ener...

Page 3: ...and other materials to prepare teeth for restoration and as an adjunct to a high speed dental handpiece and bur The system delivers a controlled mixture of abrasive particles in a dry air stream through the handpiece nozzle This manual should be thoroughly reviewed before installing and or using the equipment Please call Welch Allyn Kreativ or their representative if you need additional informatio...

Page 4: ...ded with your unit refer to your remote control User s Guide for connection Note A remote plugged into a Mach 7 will take control from the front panel A remote plugged into a Mach 8 will function as a second operatory Note A special footswitch Y adapter cable is required for the Mach 8 see page 5 CAUTION Do not exceed 135 PSI at input Applying pressure in excess of 135 PSI will cause the relief va...

Page 5: ...prior to each patient use or when changing system settings 6 Press the footswitch to start the powder flow through the nozzle Note If the handpiece is out of the cradle during power up it will need to be replaced and then removed prior to pressing the footswitch Unit Shutdown 1 Press A i r to select Standby mode 2 Set the Power Switch to OFF Note If the handpiece is left out of the cradle and the ...

Page 6: ... case and control areas with a soft cloth slightly moistened with a mild detergent or isopropyl alcohol Do not use any type of abrasive cleaners scouring powder or solvents on the case Case inside Excessive powder inside the equipment may be removed with a vacuum or a cloth slightly moistened with water or isopropyl alcohol Never use a moistened cloth on the Printed Circuit boards Powder Chamber C...

Page 7: ...n the pinch tube assembly with the grooves and securely tighten the thumb screws Monthly Requirements Internal Tubing 1 Inspect internal tubing for wear and air leakage 2 Listen for audible leaking sounds and look for powder accumulation internal to the unit 3 Soapy water and a small paintbrush are ideal for finding small air leaks 4 Clean up excess powder to help isolate the leak area 5 If a leak...

Page 8: ...e Ruptured handpiece hose Call local service rep Air flow from nozzle when footswitch is not depressed Loss of air pressure due to a misadjusted pinch valve regulator Put system in Standby and call local service rep Faulty footswitch Put system in Standby and call local service rep Defective pinch valve Put system in Standby and call local service rep Pinch valve block is not tight Put system in S...

Page 9: ...ir abrasive tubing hand pieces and handpiece tips are not warranted against normal wear and tear or customer damage This limited warranty applies only to the original purchaser user of the equipment The foregoing limited warranty is exclusive and in lieu of all other warranties whether written oral or implied and shall be purchaser s sole remedy and Welch Allyn Kreativ s sole liability under contr...

Page 10: ...Allyn Kreativ 5HF FOLQJ All recycling must be in compliance with relevant national laws Government authorized recycling centers must be used There are no hazardous materials in this product Device All covers and housings should be recycled steel or aluminum Plastic parts should be removed from the metal pads before recycling e g power inlet footswitch connector main switch etc All printed circuit ...