Directions for use
Cordless illumination system
13
Protection against ingress of
water, per IEC 60529 (illuminator,
auxiliary power cord, charging
station)
IPXØ
Operating conditions
Temperature
Relative humidity
Altitude
+10° C to +35° C (+50° F to +95° F)
30% to 75% noncondensing
700 to 1060 hPa
Shipping and storage conditions
Temperature
Relative humidity
Altitude
-20° C to +49° C (-4° F to +120° F)
15% to 95% noncondensing
500 hPa to 1060 hPa
Intermittent operation
Illuminator with speculum
Illuminator alone
10 minutes on,
5 minutes off between uses
2 minutes on,
5 minutes off between uses
Safety, EMC, and regulatory
compliance
CAN/CSA C22.2 No. 601.1-M90, including amendments 1 &
2
EN/IEC 60601-1
EN/IEC 60601-1-2
EN/IEC 60825-1
UL 60601-1
AS/NZS CISPR11
LED radiation output
< 6.67 mW at 400 – 700 nm wavelengths