Caution
Store the thermometer and probe covers in a dry location, free from dust
and contamination and away from direct sunlight. Keep the ambient temperature at
the storage location fairly constant and within the range of 50°F to 104°F (10°C to
40°C).
Take a temperature at the ear site
WARNING
Probe covers are single-use only. Re-use of a probe cover may result
in spread of bacteria and cross-contamination.
WARNING
Inaccurate measurement risk. Use only Braun ThermoScan probe
covers with this thermometer.
WARNING
Inaccurate measurement risk. Frequently inspect the probe window
and keep it clean, dry, and undamaged. Fingerprints, cerumen, dust, and other
contaminants reduce the transparency of the window and result in lower
temperature measurements. To protect the window, always keep the thermometer
in the accessory dock when the thermometer is not in use.
Caution
Inaccurate measurement risk. Before taking a temperature
measurement, make sure that the ear is free from obstructions and excess cerumen
build-up.
Caution
Inaccurate measurement risk. The following factors can affect ear
temperature measurements for up to 20 minutes:
•
The patient was lying on his or her ear.
•
The patient's ear was covered.
•
The patient was exposed to very hot or very cold temperatures.
•
The patient was swimming or bathing.
•
The patient was wearing a hearing aid or an ear plug.
Caution
Inaccurate measurement risk. If ear drops or other ear medications have
been placed in one ear canal, take the temperature in the untreated ear.
Note
A temperature measurement taken in the right ear might differ from a measurement
taken in the left ear. Therefore, always take the temperature in the same ear.
Note
When the monitor receives an ear temperature measurement, it displays the
measurement on the Home tab. If the Home tab already contains a temperature
measurement, the new measurement overwrites it.
To take a measurement and transfer it to the monitor:
1. Make sure that the monitor is powered on.
2. Remove the ear thermometer from the accessory dock.
3. Locate the probe cover box in the accessory dock.
4. Firmly push the probe tip into the probe cover box.
When the probe cover is in place, the thermometer turns on automatically.
5. Wait for the ready beep and three dashes to appear on the thermometer display.
6. Fit the probe snugly into the ear canal and then push and release the
Start
button.
•
If the probe is positioned correctly in the ear canal the ExacTemp light flashes. When the
thermometer detects an accurate measurement, the ExacTemp light is continuously on, a
long beep signals the end of the measurement, and the display shows the result.
Directions for use
Patient monitoring
77
Summary of Contents for Connex Vital Signs Monitor 6000 Series
Page 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Page 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Page 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Page 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Page 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Page 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Page 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Page 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Page 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Page 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Page 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Page 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Page 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Page 143: ......