Settings
Myogram Filter
The
Myogram filter
suppresses disturbances caused by strong muscle tremor. The set value will be the new upper limit
of the frequency range as soon as the
FILT
key is switched on or programmed as default when the unit is switched on.
When the filter is active the
‘FILT’
lamp on the unit is lit. Select the myogram filter settings as follows:
Myogram Filter
Entry Key Sequence
Setting
Confirm
ALT
8
2
Myogram Filter 25 Hz
Press
STOP
Key
3
Myogram Filter 35 Hz
1
Myo. Filt. ON when switching on unit
(marked on printout with +)
8
Myo. Filt. OFF when switching on unit
(marked on printout with –)
The myogram filter is switched on and off manually with the
FILT
key
.
Following an auto mode recording, the ECG is stored unfiltered, it is therefore possible to print the stored ECG either
with or without passing the myogram filter. Filter ON is indicated by the control lamp marked
‘FILTER’
When the
FILT
key is pressed again, the filter is switched off and the indicator lamp goes out. The cutoff frequency of
the myogram filter is set to either 25 or 35 Hz.
Settings
Myogram Filter
The
Myogram filter
suppresses disturbances caused by strong muscle tremor. The set value will be the new upper limit
of the frequency range as soon as the
FILT
key is switched on or programmed as default when the unit is switched on.
When the filter is active the
‘FILT’
lamp on the unit is lit. Select the myogram filter settings as follows:
Myogram Filter
Entry Key Sequence
Setting
Confirm
ALT
8
2
Myogram Filter 25 Hz
Press
STOP
Key
3
Myogram Filter 35 Hz
1
Myo. Filt. ON when switching on unit
(marked on printout with +)
8
Myo. Filt. OFF when switching on unit
(marked on printout with –)
The myogram filter is switched on and off manually with the
FILT
key
.
Following an auto mode recording, the ECG is stored unfiltered, it is therefore possible to print the stored ECG either
with or without passing the myogram filter. Filter ON is indicated by the control lamp marked
‘FILTER’
When the
FILT
key is pressed again, the filter is switched off and the indicator lamp goes out. The cutoff frequency of
the myogram filter is set to either 25 or 35 Hz.