80
Function check
EN
7.4 Energy supply
7.4.1
Power supply
Connect the device to the power supply.
The power supply is working perfectly if the green power supply LED comes on
permanently and the standby screen appears in the display.
7.4.2
Internal battery and energy failure alarm
1. If present: remove the replaceable battery (see “5.5 Alarm list” on page 56).
2. For the procedure for the function check of the power supply failure alarm, see the
chapter entitled “Alarms” on page 76.
If supply is not assumed by the battery without interruption, either this or the therapy
device is defective. In this case have the device including its internal battery checked by
an authorized specialist dealer or by Weinmann.
3. Check battery capacity (shown in the status line of the display).
If fewer than three segments are displayed in the battery symbol, the battery is not fully
charged. In this case, charge the battery by connecting the device to the power supply.
7.4.3
Replaceable battery (if present)
1. Push the replaceable battery into the device until you hear the latch engage.
The symbol for the replaceable battery appears on the right next to the internal battery
in the status line and a beep sounds.
2. Check battery capacity (shown in the status line of the display):
If fewer than three segments are displayed in the battery symbol, the battery is not fully
charged. In this case, charge the battery by connecting the device to the electricity
supply.
Replaceable batteries which have been stored for over a month must be charged up
before they are used to guarantee an accurate display.