52
Operation
EN
4. To switch on the device, press the On/Off key
briefly. The device is now in ventilation mode. The
patient menu can be accessed via the menu key.
When Auto switch-on (only with leakage ventilation) is
activated, you can also put on the patient/ventilator
interface and switch on the therapy device by taking a
breath (see “5.4 Activate/deactivate Auto switch-on
(only leakage ventilation)” on page 55). The operating
hours and the software version appear in the display for
about 3 seconds.
The device starts to pump air through the patient circuit. The display switches to the default
display.
5.2.3
Displays on screen
Ventilation parameters such as set therapy mode,
therapy pressures (CPAP pressure only in CPAP mode)
in hPa, the selected patient circuit and current
respiratory frequency in 1/min are shown in the
display.
Note:
1.01973 hPa correspond to 1 cm H
2
O.
The bar chart shows the pressure curve for inspiration
and exhalation.
The respiratory phase change display shows whether the
current respiratory phase was triggered spontaneously
by the patient (
S
) or by the machine (
T
). Depending on
respiratory phase, the display switches from left
(inspiration) to right (exhalation).
See the section entitled “ Symbols used in the display”
on page 18 for an explanation of the other symbols in
the display.
Vt
PEEP
IPAP
F
I : E
On/Off key
Bar chart