5. Checking the supply pressure alarm

6. Checking the ventilation rate

Information: In combination with the testing bag, these settings 
can cause the pAW /apnea alarm to be triggered. The alarm can be 
ignored during this test step.

 The number of mechanical breaths within one minute is between 

23 breaths per minute and 27 breaths per minute.

Select the following settings:
Via the adjusting knob: 
Vt = 65 ml | Freq. = 25 bpm
Via pMax button with control LEDs: 
pMax = 45 mbar

 After the oxygen cylinder has been closed, the visual and audible 

alarm for < 2.7 bar O

2

 is triggered once the device supply pressure falls 

below 2.7 bar O

2

.

 The < 2.7 bar O

2

 alarm stops when there is sufficient supply pressure.

OPEN

CLOSE

1:00 min

23-27 bpm

Summary of Contents for MEDUMAT EasyCPR

Page 1: ...hensive information please refer to the instructions for use of the device The instructions for use are available on the Internet at www weinmann emergency com Perform the function check prior to ever...

Page 2: ...accessories which have successfully passed the function check Only operate the device and accessories if they are externally undamaged Have defective devices repaired Have defective accessories repai...

Page 3: ...5 Checking the supply pressure alarm 6 Checking the ventilation rate 7 Checking the tidal volume and pAW alarm 1 Preparing for the function check 2 Visually checking the device and components 3 Checki...

Page 4: ...pressure display pMax button with LEDs for setting of max ventilation pressure Connection for ventilation hose Compressed gas connection Connection for pressure measuring tube MEDUMAT EasyCPR Overvie...

Page 5: ...Alarm mute button with LED Adjusting knob for ventilation settings Battery compartment Manual button with LED for activation of manual mode On Off button Connection for MEDUtrigger Alarm field...

Page 6: ...ice and accessories do not show any signs of damage deformations or soiling The device is connected to the oxygen supply The ventilation hose pressure measuring tube and MEDUtrigger are connected to t...

Page 7: ...y for one minute Once the device is switched on all the LEDs flash at least once and the lowest LED on the ventilation pressure display lights up green OPEN CLOSE 4 Checking visual and audio alarm out...

Page 8: ...ths within one minute is between 23 breaths per minute and 27 breaths per minute Select the following settings Via the adjusting knob Vt 65 ml Freq 25 bpm Via pMax button with control LEDs pMax 45 mba...

Page 9: ...45 mbar The device triggers the pAW alarm after the third mechanical breath at 40 mbar 50 mbar without simulated expiration After the manually simulated expiration depress the testing bag with your h...

Page 10: ...e pAW apnea alarm after the second mechanical breath 9 Checking MEDUtrigger Select the following settings Via the adjusting knob Vt 950 ml Freq 10 bpm Via pMax button with control LEDs pMax 45 mbar Th...

Page 11: ...mand flow mode The demand flow LED lights up MAT EasyCPR After the testing bag has been pressed and released a clacking sound is heard The function check is passed successfully when all of the steps h...

Page 12: ...r for Production Logistics Service WEINMANN Emergency Medical Technology GmbH Co KG Siebenst cken 14 24558 Henstedt Ulzburg Germany China Weinmann Shanghai Medical Device Trading Co Ltd T 86 21 52 30...

Reviews: