5. Checking the supply pressure alarm
6. Checking the ventilation rate
Information: In combination with the testing bag, these settings
can cause the pAW /apnea alarm to be triggered. The alarm can be
ignored during this test step.
The number of mechanical breaths within one minute is between
23 breaths per minute and 27 breaths per minute.
Select the following settings:
Via the adjusting knob:
Vt = 65 ml | Freq. = 25 bpm
Via pMax button with control LEDs:
pMax = 45 mbar
After the oxygen cylinder has been closed, the visual and audible
alarm for < 2.7 bar O
2
is triggered once the device supply pressure falls
below 2.7 bar O
2
.
The < 2.7 bar O
2
alarm stops when there is sufficient supply pressure.
OPEN
CLOSE
1:00 min
23-27 bpm