Operation
23
EN
5.5 Set SOMNO
click
300
Using a respiratory humidifier prevents the upper airways of the patient drying out during
therapy.
Six humidifier stages can be set (1 = low heating output to 6 = maximum heating output).
The most favorable setting for you will depend on room temperature and humidity, so the
ideal setting will vary with the season and ambient conditions. Preset humidifier stage 3 is
adequate under "normal" conditions.
If the temperature is 23 °C and humidity is 70 %, a humidity of 100 % can be achieved at
humidifier stage 6 and a flow rate of 20 l/min.
Switch on respiratory humidifier
1. Fill and adapt the humidifier as described in the associated instructions for use.
2. Switch on the therapy device.
3. Press the humidifier key on the therapy device.
The humidifier symbol
and the humidifier stage
(Stage 3 in this case) are shown in the display of the
therapy device.
Set humidifier stage
1. Call up the patient menu and scroll to the setting for humidifier stage.
or:
with the device switched on: keep the humidifier key depressed until the current
setting appears.
2. To change the heating stage, press the Softstart key (+) or the humidifier key (-) several
times until the desired stage is shown.
If you have dry airways in the morning, heating output is set
too low
. Select a higher
setting in this case.
If condensation forms in the breathing tube during the therapy night, heating output
is set
too high
. In this case, select a lower humidifier stage.
3. To save the setting and exit the menu, press the on/off key/ESC or wait until the menu
exits automatically.