42
Repairguide for VENTImotion/VENTIlogic
7.9 Replace battery on control board
Tools and equipment required
• ESD protected work-station
• 3 mm Allen key
• Phillips screwdriver, size 1
1. Open the device (see "7.2 Open device" on
2. Undo the screw
on the battery holder and
take spring
Put the parts on one side.
3. Take battery
out of the holder on the control
board.
4. Put in a fresh battery
5. Insert spring
6. Close the device (see "7.3 Close the device"
7.
Reset the date and time
as described in the
VENTI
motion
/VENTI
logic
hospital manual.
Caution!
After changing the control board, always set the
date in the following sequence: 1. YEAR,
2. MONTH, 3. DAY, otherwise you may get a
faulty display.
8. Test the device (see "4. Test the device" on
7.10 Dismantle device outlet
Tools and equipment required
• ESD protected work-station
• 3 mm Allen key
• Phillips screwdriver, size 1
1. Open the device (see "7.2 Open device" on
2. Remove the mains supply unit (see "7.4 Re-
place mains supply unit" on page 33).