Repairguide for VENTImotion/VENTIlogic
31
7. Repairguide for VENTI
motion
/VENTI
logic
7.1 General
Perform repairs on the VENTI
motion
or VENTI
logic
exclusively at an ESD protected work-station!
•
Follow the safety rules in the operating instruc-
tions for the therapy device.
• Any handling of the device assumes accurate
knowledge and observance of the operatingin-
structions and the service and repair guide.
• Only perform repairs which are described in
this service and repair guide. This is the only
way to guarantee that the therapy device con-
tinues to function properly.
• Ensure that your hands and your work-station
are clean during repair work.
• Perform a function check after
every
(see "4. Test the device" on page 13).
• If you replace components or individual parts,
use only original Weinmann parts.
• When ordering the bottom part of the
housing
, quote the type, year of manufac-
ture and number of the device as well.
Note
The item numbers listed in the text which fol-
lows are identical to the item numbers in the
spare parts list on page 48 and in the over-
view on page 4.
7.2 Open device
Tools and equipment required
• ESD protected work-station
• 3 mm Allen key
Caution!
Open the device only with the power supply plug disconnected.
1. Place the device on a non-slip surface with the
top facing downwards.
2. Remove the filter cassette from the top part of
the housing.
Summary of Contents for BiLevel-ST VENTImotion WM 24800