20

Hygienic preparation

3. Push the free end of the intermediate tube 

20

 onto 

the aspiration tube port 

19

. This closes the dis-

posable collection bag 

21

 and thereby reduces 

the risk of secretion escaping.

4. Remove the disposable collection bag 

21

 and dis-

pose of it together with the aspiration tube 

19

.

Important

Be sure to observe the relevant rules when dispos-
ing of the collection bag (see “2. Safety” on 
page 5).

5. Remove the vacuum tube 

22

.

6. Lift the disposable canister 

7

 from its holder 

25

.

7. Remove the tube holder plate 

9

 by pushing it to the 

left until it disengages and then pulling it towards 
you.

5.2  Cleaning, disinfecting and sterilizing

Hygienic preparation of the ACCUVAC Basic and the 
accessories used should be performed as described in 
the following table.

Observe the instructions regarding use of disinfectant. For 
immersion disinfection we recommend GIGASEPT FF  
and TERRALIN for wipe disinfection. 

Summary of Contents for ACCUVAC Basic

Page 1: ...Description and Operating Instructions ACCUVAC Basic Aspirator WM 10700 WM 10709...

Page 2: ...spiration 16 4 4 Charging ACCUVAC Basic 16 5 Hygienic preparation 19 5 1 Preparations 19 5 2 Cleaning disinfecting and sterilizing 20 5 3 Reassembling the disposable collection canister 22 6 Functiona...

Page 3: ......

Page 4: ...ter 2 Vacuum control 3 Voltage indicator 8 Motor unit 6 Release catch 4 On Off switch 9 Tube holder plate 1 Pressure gage 5 Membrane keyboard ACCUVAC Basic from rear without collection canister 13 Muf...

Page 5: ...5 Holder 11 Adapter charger plug 22 Vacuum tube Accessories 38 Holder set WM 15271 Connection cable WM 10650 Accessory bag WM 10655 Wall bracket WM 15208 13 8 V Mains charger unit WM 2645 27 Vent tab...

Page 6: ...plints ACCUVAC Basic can when used by a skilled operator eliminate obstruction of the respiratory tract and hence the risk of respiratory failure cuts energy consumption by reducing power output on re...

Page 7: ...interme diate tube 20 the T piece 23 and the vacuum tube 22 The disposable collection canister 7 is inserted in the holder 25 screwed on the side of the motor unit 8 The vacuum tube 22 of the disposab...

Page 8: ...ble as an accessory Important The AC adapter charger plug 11 supplied with the appliance is only designed for charging the power pack with the appliance switched off It must not be used to operate the...

Page 9: ...points Please read these operating instructions carefully They belong to the appliance and must be kept available at all times Before working with ACCUVAC Basic you must understand how to use it Be s...

Page 10: ...ctions for use or original spare parts are not utilised During the aspiration process take particular care not to cause injuries to the patient s mouth or throat e g to mucous membrane The suction can...

Page 11: ...r pack first as described in 4 4 Charging ACCUVAC Basic on page 16 Lead cell power packs discharge even during non use The rate of auto discharge increases with ris ing temperature and amounts to abou...

Page 12: ...late shows the space required 2 Hold up the drilling template and level it with a spirit level 3 Mark the required screw holes on the mounting location 4 Drill the screw holes with a 3 mm drill 5 Scre...

Page 13: ...is supplied complete with a red adapter ring This is needed if you want to plug the vehicle plug into a cigarette lighter socket To connect to the mains charger unit WM 2645 accessory or a 12 volt ve...

Page 14: ...ion kit is available under order number WM 15261 1 Remove the disposable collection canister 7 from its holder 25 2 Unscrew the holder 25 from the motor unit 8 3 In the same place screw on the holder...

Page 15: ...ge 11 1 Use the velcro type strips to attach the accessory bag to the loops on the motor unit 3 4 Fitting a rinsing glass On the left side of the appliance it is possible to fit an additional collecti...

Page 16: ...12 Assembly 2 Fix the holder 38 to the left side of the appliance 3 Push the rinsing glass into the holder...

Page 17: ...up Do not switch on ACCUVAC Basic while there is still a vacuum This would cause the pump motor to draw too much current when switching on which could actuate fuse F2 16 4 Set the desired vacuum at th...

Page 18: ...ter inte grated in the disposable collection bag 21 is not wet ted by secretion If the overflow valve filter is wetted with secretion it becomes impermeable and the dis posable collection bag 21 has t...

Page 19: ...collection canister 24 3 Push the free end of the intermediate tube 20 onto the connection for the aspiration tube This closes the disposable collection bag 21 and reduces the risk of fluid escaping 4...

Page 20: ...ACCUVAC Basic as soon as the yellow voltage indicator LED lights up By doing so you will ensure that it operates for an adequate period next time it is used The power supply for charging must deliver...

Page 21: ...an damage ACCUVAC Basic Charging without wall bracket 1 Switch off ACCUVAC Basic 2 Place the plug 11 at the appliance end of the con necting cable in the guide slot at the rear of the ap pliance and p...

Page 22: ...e the bat tery for another two 2 hours at least Refresher charging Regular refresher charging extends the service life of the power pack Leave ACCUVAC Basic connected to the external power supply even...

Page 23: ...Never immerse the ACCUVAC Basic motor unit in dis infectant or other liquids Always disinfect simply by wiping with disinfectant Otherwise you may damage the device and thereby endanger users and pati...

Page 24: ...dispos ing of the collection bag see 2 Safety on page 5 5 Remove the vacuum tube 22 6 Lift the disposable canister 7 from its holder 25 7 Remove the tube holder plate 9 by pushing it to the left until...

Page 25: ...osable collection canister Collection canister In warm water with mild house hold cleaner Immerse in dilute solution 3 Rinse at up to 95 C Hot steam sterilization up to 121 C 4 Vacuum tube Not permitt...

Page 26: ...liding it to the right 2 Insert the collection canister 7 in its holder 25 3 Use the vacuum tube 22 to connect the T piece of the collection canister 24 to the suction port 14 on the motor unit 8 4 In...

Page 27: ...on bag 21 WM 10732 Aspiration tube with fingertip 19 WM 10733 6 1 Intervals To ensure that a properly functioning ACCUVAC Basic is always available it is essential to observe the following intervals B...

Page 28: ...ging ACCUVAC Basic on page 16 At least every 6 months Perform a functional check see 6 2 Performing the functional check on page 25 Make a visual inspection of the muffler for clog ging If it is clogg...

Page 29: ...lways recharge the power pack without delay if the yellow LED on the voltage indicator lights up see 4 4 Charging ACCUVAC Basic on page 16 5 Insert the stopper in the fingertip 6 Use your thumb to hol...

Page 30: ...rly engaged Make sure connection snaps in Appliance does not switch on or off Electronic fault Have it repaired Appliance does not reach maximum vacuum of 0 8 bar in 20 seconds but voltage indicator s...

Page 31: ...charger plug WM 10750 or charger WM 2645 accessory Vehicle electrical system is not supplying 12 0 13 8 V Check vehicle electrical system Power pack damaged by being fully discharged Fit new power pa...

Page 32: ...nnect the power pack connector 18 X1 from the circuit board 7 Unscrew the power pack holder 4 cross head screws 8 Detach both leads from the power pack 17 9 Remove the faulty power pack Help protect t...

Page 33: ...ther again making sure that the silicone sealing cord is correctly inserted and is not jammed or otherwise damaged 14 Perform a functional check see 6 2 Performing the functional check on page 25 7 2...

Page 34: ...ntral contact of the plug is the plus pole The plus lead of the cable has either a square cross section or colored markings The outer contact of the plug is the minus pole The minus lead of the cable...

Page 35: ...the cover plate 2 cross head screws 2 Remove the old muffler 13 3 Insert a new muffler 4 Refit the cover plate Make sure that the pad stuck to the cover plate is pressing against the muffler 5 Perform...

Page 36: ...e recommend that you have any servicing such as inspections and repair work carried out by the manu facturer Weinmann or by expert personnel 8 2 Disposal Do not dispose of the unit with domestic waste...

Page 37: ...er ACCUVAC Basic WM 10700 consisting of Aspirator unit disposable canister WM 10710 AC adapter charger plug complete WM 10750 2 ACCUVAC Basic with re usable canister ACCUVAC Basic WM 10709 consisting...

Page 38: ...M 15261 consisting of Set re usable canister WM 15269 Holder set WM 15271 6 Set re usable canister WM 15269 consisting of Re usable canister WM 10630 Instructions for use WM 16235 7 Holder set for re...

Page 39: ...WM 15315 Set III 30 x WM 5156 WM 15316 11 Aspiration catheter sterile pack CH 10 diam 3 3 mm 50 cm long pack of 4 WM 5157 Set I 10 x WM 5158 WM 15317 Set II 20 x WM 5158 WM 15318 Set III 30 x WM 5158...

Page 40: ...der plate green WM 10723 5 Lead cell power pack WM 10747 6 Muffler WM 10665 7 Instructions for use WM 16138 8 Fuse for vehicle plug WM 10673 9 Fuse for external power supply F1 WM 2692 10 Fuse for pow...

Page 41: ...ctric shock BF Degree of protection against water IPX 1 drip water Classification according to EN ISO 10079 1 high vacuum high flow Norms complied with EN 60601 1 EN ISO 10079 1 EN 1789 Vehicle plug f...

Page 42: ...e high fre quency telecommunication devices transmitters and the ACCUVAC Basic depending on the rated output of the transmitter as given below Recommended safe distances between portable or mobile HF...

Page 43: ...repair Acts of God e g lightning strikes etc Transport damage as a result of improper packaging of returned items Lack of maintenance Operational and normal wear and tear which inclu des for example t...

Page 44: ......

Page 45: ......

Page 46: ......

Page 47: ......

Page 48: ...ternet www weinmann de For decades Weinmann has been developing producing and marketing medical devices for markets around the world In cooperation with our partners we design economic health systems...

Reviews: