Timer delay
dL1=
0
Timer delay (set to 0 s to disable)
e.g., no delay
-
Dec
-
Inc
Next
Accept
Calibrationoptions
Calibrate
Input?
CIPn
CIPy
Skip input calibration
Calibrate inputs
Toggle
Accept
Calibrate
Output?
COPn
COPy
Skip output calibration
Calibrate outputs
Toggle
Accept
Save values
SAVE
Instrument is saving the changes to the setup and returning to normal
operation
*
Note: Alarms and Analogue outputs are fitted to the PMX420Plus only. You can still set values for the
alarms and analogue output on the PMX420 but they will only affect the display and status leds.
During setup the
Å
and
ì
keys change values and toggle options.
The
ENT
key accepts changes and moves to the next item in the setup
sequence (see table).
Note: if you remove the security link during the setup sequence, any
changes you have made will not be saved.
• InputCalibration
General
The PMX0 series instruments are factory calibrated. You can change
the input range without calibration, but the outputs are calibrated for
the specific range given during the setup procedure.
Equipmentrequirements
• A suitable current or voltage source
TerminalConnectionsforinputcalibration
CalibrationStage
Signaltype
Terminal
Current (mA) inputs
Milliamp source (+)
8
Milliamp source (−)
9
Voltage (V) inputs
Voltage source (+)
7
Voltage source (−)
9
InstrumentSetup(continued)