![Wealtec HB-1 Installation And Operation Manual Download Page 19](http://html1.mh-extra.com/html/wealtec/hb-1/hb-1_installation-and-operation-manual_4011632019.webp)
18
international standard thermometer. Some temperature variation may occur
with different thermometer.
If necessary to perform any adjustments in the field, please follow procedure as
below:
A. Put a thermometer in the hole of the aluminum block and operate Block
Heater in same way as the Operation section above to start the Block Heater.
B. The OFFSET function is activated only under running condition and the
temperature of block is reached and stabilized at the desired set value
(Temperature LED display stops flashing at this moment).
C. When press the Temperature Up or Down-key under the situation as
described above, the Temperature LED display monitor will be off and Timer
LED display will show 000.
*Note!
The function display will be de-activated within 3 seconds if no further
press on Up or Down-key. Press Temperature Up or Down-key to activate the
OFFSET function again.
D. If desired value is lower then displayed value, press down arrow button of
Temperature setting. If desired value is higher then displayed value, press up
arrow button of Temperature setting. The maximum adjustable range is +/-100