PROBLEM

POTENTIAL CAUSES

SOLUTIONS

Noise current is higher than expected.

There may be ground loops in the 

system; some system components 

may be higher noise than expected; 

cabling and shielding may be 

inadequate.

Refer to Wavelength Electronics Application 

Note 

AN-LD08: Manage Grounding to 

Minimize Noise with the QCL Drivers

:

www.teamwavelength.com/download/

applicationtechnotes/an-ld08.pdf

Also reference 

AN-LD09: Troubleshooting 

Low Noise Systems

:

www.teamwavelength.com/download/

applicationtechnotes/an-ld09.pdf

The output switched off unexpectedly

The instrument detected an internal 

error.  

(See 

Error Messages

 on 

page 19

.)

Switch the back panel power button off and 

then on again, and turn the front panel power 

button on.

The internal temperature has 

exceeded 70ºC.

Allow the instrument to cool down and then 

press the front panel power button.

20 kHz noise is in the output current

QCL driver is operating in current limit.

If safe for the quantum cascade laser, 

increase the output current limit setting.  

When actively operating in current limit, the 

safety circuits add noise to the output.

If I run my QCL at about 5% of maximum 

output current, the transfer function 

becomes non-linear and I see distortion in 

my output current waveform.

QCL driver resolution is too low.

Select a QCL driver with a much lower 

maximum output current.  For example, 

to run 50 mA, request a QCL125 (125 mA 

maximum output).

My instrument doesn’t show up in the list.  

(

Remote Only

)

Your instrument may not be turned on.

Power up the instrument and click 

Refresh

 

on the 

Choose Instrument

 window.

Your instrument may not be plugged in 

to the computer.

Check that either a USB or Ethernet cable is 

tightly connected to the instrument and to the 

computer.

Your instrument may be on a different 

sub-net.

Add the sub-net to the 

Choose Instrument 

window and then click 

Refresh

.

Actual current does not match the setpoint.

The setpoint exceeds the limit current 

setting.

If the limit current can be increased without 

damage to the QCL, increase the limit 

current setting.

The QCL is compliance voltage 

limited.

Increase the Max Supply Voltage slightly to 

see if actual current increases.  Follow the 

instructions on 

page 20

 to optimally set 

the Max Supply Voltage.

Summary of Contents for QCL500 LAB

Page 1: ...t RCKMT LABDUAL 19 Double Unit Rack Mount Kit INTLK REPL KIT included with purchase Interlock Replacement Kit with BNC terminator and modified D SUB connector to override interlocks e Pb RoHS Complian...

Page 2: ...ailable Powering your QCL with this patented1 driver gives you narrow linewidth stable center wavelength and repeatable scans INTUITIVE USER INTERFACE AND SUPERIOR SOFTWARE CONTROL With Wavelength s p...

Page 3: ...ing power To avoid fire hazard use only the specified fuse with the correct type number voltage and current ratings and use only the recommended replacement parts CAUTIONS The QCL instrument has no us...

Page 4: ...Then turn the drawer until the correct voltage label is at the top and replace it in the receptacle Press firmly to make sure the drawer is completely seated It should be flush with the receptacle Fig...

Page 5: ...blue Limit Current value and rotate the Adjustment knob to change to the desired value The instrument will beep when set NOTE The Max Supply Voltage and Cable Resistance values can be ignored for now...

Page 6: ...can source 250 mA 2 Alignment LED ground 3 QCL current enabled indicator 5 V TTL high disabled low ground enabled 4 QCL current enabled indicator ground 5 Not used 6 Analog setpoint monitor 7 Analog...

Page 7: ...XTERNAL MODULATION MIN TYP MAX UNITS NOTES Modulation Bandwidth 2 3 MHz Sinewave Rise Fall Time 250 200 nsec To full scale Slew Rate 30 V sec Depth of Modulation 90 At 500 kHz External Modulation Inpu...

Page 8: ...atts At idle no output to QCL load GENERAL SPECIFICATIONS VALUE NOTES Warm up Time 1 Hour Remote Interface USB Ethernet Warranty One 1 Year Safety Tested To IEC 61000 3 2 IEC 61000 3 3 IEC 61000 4 2 I...

Page 9: ...rive current for the laser Over and under voltage protection circuits protect the driver and QCL if the power supply voltages fall outside of the acceptable operating range Over temperature protection...

Page 10: ...cal earth ground 3 QCL SMA SMA connector alternate connection to QCL 4 QCL D SUB 9 Pin female D SUB connector 5 Monitors 9 Pin male D SUB connector 6 USB USB port connection between instrument and a c...

Page 11: ...receptacle Press firmly to make sure the drawer is completely seated It should be flush with the receptacle Figure 12 Fuse Installation and Voltage Selector 5 Close the faceplate again pressing firml...

Page 12: ...ument beeps and text returns to blue when values are set Touch again to cancel selection ICONS Icon color indicates current state Blue Screen is active Grey Screen is inactive touch icon to access tha...

Page 13: ...aximum Internal Supply Voltage can be lowered to minimize internal instrument temperature or to act as a compliance voltage clamp on the QCL output current See Additional Technical Notes on page 20 fo...

Page 14: ...n press the red Stop button to abort the scan Any data collected during the aborted scan is saved to the USB flash drive NOTES The step interval time is approximate In the scan files time intervals ar...

Page 15: ...ernal modulation signal voltage see the next section for the transfer function and use the following equation VEXT_MOD ISETPOINT Transfer Function If the external modulation input causes the driver to...

Page 16: ...TTL indicator If current to the QCL output is enabled the voltage between PIN 3 and PIN 4 on the MONITORS D SUB will be low It will be high if current is disabled USE THE TRIGGER SIGNAL A trigger sig...

Page 17: ...rument experienced an internal fault See Error Messages on page 19 Switch the back panel power button off and then on again and turn the front panel power button on The internal temperature has reache...

Page 18: ...hen actively operating in current limit the safety circuits add noise to the output If I run my QCL at about 5 of maximum output current the transfer function becomes non linear and I see distortion i...

Page 19: ...instrument cannot drive current through the load Make sure the QCL SMA or Pins 5 9 of the QCL D SUB are connected A short circuit was detected at the load terminals Please verify the load and load co...

Page 20: ...ge input and the current is still not compliance voltage limited the instrument is operating at its lowest voltage and the voltage limit cannot be reduced further USE PASSIVE AND ACTIVE INTERLOCKS The...

Page 21: ...found at www teamwavelength com support software downloads 3 Extract the file on the flash drive It should contain the following files INSTALL INI QCLxx ABL IMG QCLxx APP IMG QCLxx INSTALL INI NOTE Q...

Page 22: ...22 MECHANICAL SPECIFICATIONS 0 72 3 89 0 66 16 8 3 92 99 6 6 10 154 9 8 49 215 6 Dimensions are in inches mm Tolerance is 5 8 860 225 0 3 47 88 14 8 860 225 0 16 35 415 16 35 415 17 72 450 3 47 88 14...

Page 23: ...patibility tested to IEC 61000 3 2 and IEC 61000 3 3 IEC 61000 4 2 IEC 61000 4 3 IEC 61000 4 4 IEC 61000 4 5 IEC 61000 4 6 and IEC 61000 4 8 European Council Directive 2014 35 EC 26 February 2014 on L...

Page 24: ...tion contained in this document is subject to change without notice Wavelength will not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishi...

Reviews: