Maintenance
UV Lamps have a
1 year (9000 Hour) life span
under normal
operating conditions.
Quartz sleeves should be cleaned with vinegar, citric acid, or a
lime scale removing chemical annually and replaced no less than
once every 3 years. If the quartz sleeve can not be cleaned, or if
it is discolored, is must be replaced.
Pre filters should be maintained according to the manufactur-
ers instructions to ensure feed water requirements within this
manual are met.
Lamp and Quartz Sleeve Replacement Procedure
1. Disconnect the power supply from the wall outlet.
2. Turn off the inlet valve and outlet isolation valves.
3. Disconnect the lamp lead wire from the glow cap electrical
port. # 1
4. Rotate the glow cap counter clockwise to unlock it from the
mounting bracket. # 2
5. Pull up on the glow cap to remove the lamp from the quartz
sleeve #3. If not replacing the quartz sleeve proceed to step
#11.
6. De-pressurize the system.
7. Grip the quartz sleeve nut with your hand and unscrew coun-
ter clockwise until it can be lifted from the reactor chamber's
threaded port. # 4
8. Remove the quartz sleeve from the reactor chamber. # 5
9. Fully insert a new quartz sleeve into the reactor chamber
and install the o-ring over the quartz sleeve. Press the o-ring
down over the quartz sleeve until it is seated in the bevel of
the stainless steel reactor chamber's threaded port.
10. Install the quartz sleeve nut onto the threaded port of the
reactor chamber. Screw the nut on clockwise and tighten
hand tight.
11. Insert a new lamp into the quartz sleeve.
12. Gently press down on the glow cap and rotate it clockwise
until it locks in place on the mounting bracket
13. Reconnect the lamp lead wire to the new lamp.
14. Turn on the inlet valve and check for leaks. Repair leaks as
needed.
15. Restore power to the system.
16. Verify new bulb has 365 days remaining on the controller
screen.
Immediately after replacement of the UV lamp, and/or quartz
sleeve, follow the disinfection procedure below.
NOTICE
If any water comes in contact with any electrical components,
dry the components immediately.
DO NOT
energize a system
with damp or wet electrical components.
WARNING
!
DO
NOT
touch the quartz sleeve with your fingers. Hold it with a
paper towel.
DO
NOT
touch the lamp with your fingers. Hold it with a paper
towel.
WARNING
!
WARNING
!