background image

IS-HF-M100+-29293  Rev 5

 

8

 

Calibrating the M100+

The instrument was calibrated and tested prior to leaving the factory, so 

it is possible to use the instrument directly out of the box. Under normal 

conditions, recalibration is required once every three months. Quarterly 

calibration ensures performance within accuracy specifications.
The EPA and ISO recommend that on-line turbidimeters used for 

reporting purposes be calibrated with a primary standard at least once 

every three months.

Calibration Standards

If the M100+ will be used over the entire range of 0.02 to 4000 NTU, 

a complete calibration as described below will be required, which 

includes two High Range standards.
If the instrument accuracy is only required below 1000 NTU, a 

calibration may be performed using the 1000 NTU standard, the 10 

NTU standard and the 0.02 NTU standard. To calibrate starting at 1000 

NTU, simply press the 1000 button on the calibration screen.
If the instrument accuracy is only required below 10 NTU, such as for 

potable water, a calibration may be performed using only a 10 NTU and 

a 0.02 NTU standard. To calibrate starting at the 10 NTU, simply press 

the 10 button on the calibration screen.
We recommend that the following materials be used during calibration 

to achieve the full-scale accuracy stated in this manual. All calibration 

standards listed below are available for purchase from HF scientific.

•  High Range A ProCal Calibration Standard
•  High Range B ProCal Calibration Standard
•  1000 NTU ProCal Calibration Standard
•  10.0 NTU ProCal Calibration Standard
•  0.02 NTU ProCal Calibration Standard

Diluted Formazin is unstable. If Formazin is used to calibrate the 

instrument, ensure that a fresh stock suspension of Formazin 

is used to achieve the accuracy quoted for the instrument. A 

Formazin Stock Solution Kit (catalog no. 50040) is available from 

HF scientific. 

NOTICE

The HF scientific ProCal (Primary Calibration Standards) are more 

stable than Formazin and have a minimum shelf life of 12 months. For 

more information on ordering HF scientific's ProCal Standards, see 

Accessories and Replacement Parts List later in this manual. 
Prior to recalibration, review the expiration dates to ensure that the 

standards have not expired.

Care of ProCal Standards

The information below is provided as general guidelines for proper care 

of ProCal standards.

•  Standards should only be handled by the top, black cap or by 

the very bottom of the cuvette.

•  Keep cuvette glass clean of fingerprints and debris.
•  Cuvettes can be cleaned with any domestic glass cleaner and 

the provided microfiber cloth.

•  Prevent ProCal standards from freezing. This will most likely 

destroy them. 

•  Standards under 100 NTU must be freshly poured from the 

supplied bottle as they are not stable in the glass cuvette.

•  There is no need to shake standards, but standards over 1000 

NTU should be up-ended gently before use.

•  Replace standards at the expiration date.

Indexing Calibration Cuvettes

To achieve the greatest accuracy and account for normal scratches 

and aberrations in cuvette glass when calibrating, HF scientific 

recommends indexing the cuvettes.
Standards and standards kits purchased from HF scientific are supplied 

with indexing rings. Complete instructions on how to index cuvettes are 

included in the calibration kits. 
Indexing is best performed if the automatic reading is enabled and 

set to 1 sec/reading. See Auto/Manual Readings section for more 

information. The following steps allow repeatable indexing of calibration 

standards.
1.  With the instrument in the home screen, insert the standard.
2.  Slowly rotate the standard inside the optical well one complete turn 

(360°). 

3.  While rotating the standard slowing, observe the measured turbidity 

and locate the position of the cuvette having the lowest reading.

4.  With the calibration standard positioned at the location having the 

lowest turbidity reading, install the indexing ring over the cap on 

the standard so that the pointer of the indexing ring faces directly 

forward.

When using the standards in the future, always insert the standard so 

that the pointer of the indexing ring faces forward. Slowing rotate the 

standard back and forth about 5° to find the lowest point. The standard 

is now indexed and ready for use.

Calibration Procedures

To calibrate for 0-4000 NTU using ProCal, you will need the following 

standards:

•  High Range A
•  High Range B
•  1000 NTU
•  10 NTU
•  0.02 NTU standards.

To calibrate for 0-4000 NTU using Formazin, you will need the following 

standards:

•  4000 NTU
•  1000 NTU
•  10 NTU
•  0.02 NTU

Prior to beginning calibration, clean the calibration standard cuvettes 

with a microfiber cloth and standard window cleaner. Take care to only 

handle standards by the black cap to prevent fingerprints or smudges 

from affecting the readings.

Summary of Contents for HF scientific M100 28060

Page 1: ...een Calibration 6 Set Date Time 6 Delete Samples 6 Data Resolution 6 Offset 7 Device ID 7 Calibration Reminder 7 Calibration Download 7 Firmware Update 7 Reset Defaults 7 About Device 7 Calibrating th...

Page 2: ...ould not permit anyone to touch this equipment unless they are over 18 years of age are adequately trained and supervised and have read and understood this manual Owners should also ensure that no una...

Page 3: ...t in death or serious injury This symbol identifies hazards which if not avoided could result in death or serious injury This symbol identifies hazards which if not avoided could result in minor or mo...

Page 4: ...ns for the M100 are valid only for measurement of static non flowing samples Touch Screen Standard Inserted in Optics USB Connector Unpacking and Inspection of the Instrument Remove all items from pac...

Page 5: ...gned to provide the ability to customize the instrument according to your needs at any time during normal operation The Setup screen has been split into sub menus to facilitate instrument configuratio...

Page 6: ...ation The screen has been factory adjusted but can be adjusted if the alignment is off The screen may need to be calibrated if new software has been downloaded It is a 5 position setup 1 Press the Set...

Page 7: ...sert the USB memory stick that contains the software into the USB port 3 If the firmware revision on the memory stick is older than the firmware currently installed the instrument will ask if you wish...

Page 8: ...ded as general guidelines for proper care of ProCal standards Standards should only be handled by the top black cap or by the very bottom of the cuvette Keep cuvette glass clean of fingerprints and de...

Page 9: ...n on the left side of the screen 2 Insert the 4000 NTU standard 3 Press the Command button in the middle of the screen to start calibration A countdown from 30 will appear on the screen When the count...

Page 10: ...stic Banner On Off Reading List Setup Calibration Trend Home Data Retrieval Data logging can only save 100 points per Sample ID All data is saved in CSV format and is date and time stamped Sample ID D...

Page 11: ...Email HF Info wattswater com www hfscientific com Routine Maintenance Routine maintenance should include calibration of the instrument See Calibration section in this manual for more information Clea...

Page 12: ...al or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting...

Reviews: