background image

Installation Instructions
M100+

TM

 Laboratory Turbidimeter

White Light - Model 28060
Infrared - Model 28061

IS-HF-M100+-29293  Rev 5  2137

Table of Contents

Attention Owners and Users .........................................................................................2

Product Identification ..............................................................................................2

Understanding Safety Information ................................................................................3
Introduction to the Unit ................................................................................................4

Overview .................................................................................................................4
Specifications .........................................................................................................4
Unpacking and Inspection of the Instrument ............................................................4

Site Selection ...............................................................................................................4

Power Requirements ...............................................................................................4

Setup ...........................................................................................................................4

Operating Screens ..................................................................................................5

Home Screen......................................................................................................5
Trend Screen ......................................................................................................5
Calibration Screen ..............................................................................................5
Setup Screen ......................................................................................................5

Configuring the Instrument ......................................................................................5

Turning on the Instrument...................................................................................5
Screen Brightness ..............................................................................................5
Auto/Manual Readings........................................................................................6
Auto Power Down ...............................................................................................6
Touch Screen Calibration ....................................................................................6
Set Date/Time ....................................................................................................6
Delete Samples ..................................................................................................6
Data Resolution ..................................................................................................6
Offset .................................................................................................................7
Device ID ............................................................................................................7
Calibration Reminder ..........................................................................................7
Calibration Download .........................................................................................7
Firmware Update ................................................................................................7
Reset Defaults ....................................................................................................7
About Device ......................................................................................................7

Calibrating the M100+ ............................................................................................8

Calibration Standards .........................................................................................8
Care of ProCal Standards ....................................................................................8
Indexing Calibration Cuvettes .............................................................................8
Calibration Procedures .......................................................................................8
Calibration Procedures (continued) .....................................................................9
Calibrating using ProCal Standards .....................................................................9
Calibrating Using Formazin Standards ................................................................9
Alternate Calibration Ranges ..............................................................................9
Failed Calibration ................................................................................................9

Operation ...................................................................................................................10

Home Screen ........................................................................................................10
Trend Screen .........................................................................................................10
Data Retrieval .......................................................................................................10

USB Connection ................................................................................................10

Troubleshooting .........................................................................................................11

M100+ Fault Detection .........................................................................................11
Diagnostic Table ....................................................................................................11
Technical and Customer Assistance .......................................................................11

Routine Maintenance .................................................................................................11

Cleaning the Cuvette .............................................................................................11
Replacing the Source Lamp...................................................................................11
Battery ..................................................................................................................11

Accessories and Replacement Parts ...........................................................................11
Warranty ....................................................................................................................12

Summary of Contents for HF scientific M100 28060

Page 1: ...een Calibration 6 Set Date Time 6 Delete Samples 6 Data Resolution 6 Offset 7 Device ID 7 Calibration Reminder 7 Calibration Download 7 Firmware Update 7 Reset Defaults 7 About Device 7 Calibrating th...

Page 2: ...ould not permit anyone to touch this equipment unless they are over 18 years of age are adequately trained and supervised and have read and understood this manual Owners should also ensure that no una...

Page 3: ...t in death or serious injury This symbol identifies hazards which if not avoided could result in death or serious injury This symbol identifies hazards which if not avoided could result in minor or mo...

Page 4: ...ns for the M100 are valid only for measurement of static non flowing samples Touch Screen Standard Inserted in Optics USB Connector Unpacking and Inspection of the Instrument Remove all items from pac...

Page 5: ...gned to provide the ability to customize the instrument according to your needs at any time during normal operation The Setup screen has been split into sub menus to facilitate instrument configuratio...

Page 6: ...ation The screen has been factory adjusted but can be adjusted if the alignment is off The screen may need to be calibrated if new software has been downloaded It is a 5 position setup 1 Press the Set...

Page 7: ...sert the USB memory stick that contains the software into the USB port 3 If the firmware revision on the memory stick is older than the firmware currently installed the instrument will ask if you wish...

Page 8: ...ded as general guidelines for proper care of ProCal standards Standards should only be handled by the top black cap or by the very bottom of the cuvette Keep cuvette glass clean of fingerprints and de...

Page 9: ...n on the left side of the screen 2 Insert the 4000 NTU standard 3 Press the Command button in the middle of the screen to start calibration A countdown from 30 will appear on the screen When the count...

Page 10: ...stic Banner On Off Reading List Setup Calibration Trend Home Data Retrieval Data logging can only save 100 points per Sample ID All data is saved in CSV format and is date and time stamped Sample ID D...

Page 11: ...Email HF Info wattswater com www hfscientific com Routine Maintenance Routine maintenance should include calibration of the instrument See Calibration section in this manual for more information Clea...

Page 12: ...al or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting...

Reviews: