Restore Factory Settings.
6.1 Restoring Factory Settings
If the instrument is unable to perform a calibration due to a low lamp output, bad
calibration standard or a dirty cuvette, the instrument will display
CAL
on the lower row
of the display and
Err
on the upper row. The operator has two choices to correct this
problem or restore the factory calibration. If the operator can determine whether a poor
calibration or a low lamp caused the problem, he/she can remedy the problem and
recalibrate. If all else fails, the operator may restore the factory calibration and
configuration settings by performing the following operation. Push and hold the
button.
Now push and release the
↵
then release the
button. Factory calibration and factory
configuration have now been restored.
Restoring the factory settings allows the use of the AccUView LED with reduced
accuracy. The original problem still exists and must be determined and corrected
before accurate operation of the AccUView LED will be resumed. The AccUView
LED will always reset to %T and factory defaults after restoring.