Watson-Marlow 720UN and 720SN User Manual
44
16.6 Auto-restart
The pump offers an auto-restart feature. If active on power loss, it restores the pump
when power returns to the operating state it was in when power was lost. It does not
operate when powering down in the middle of a dose: when the pump is restarted, it
will await a press on the
START
key to begin the interrupted dose again. Auto-restart
is retained while the pump is switched off. When the pump starts running, look for the
!
symbol on the display. This symbol indicates that the pump is set for auto-restart.
In the second screen of the Setup menu select
Auto-restart
using the
UP
and
DOWN
keys. Press
ENTER
to confirm your choice.
The pump displays a screen allowing the user to activate auto-restart. Use the
UP
and
DOWN
keys to choose
On
or
Off
and press
ENTER
to confirm the deci-
sion.
If
Off
is chosen, the pump returns the user to the second screen of the Setup
menu. The auto-restart facility will not operate.
If
On
is chosen, the pump returns the user to the second screen of the Setup
menu, where an exclamation mark (
!
) is now visible. This mark confirms that
the auto-restart feature is in place and will operate the next time power is lost
and restored.
Alternatively ...
Stop the pump. Turn off the mains power switch at the rear of the pump.
Hold down the
START
key and turn on the mains power switch. The
!
symbol
shows on the display.
Start the pump. If the mains supply is interrupted the pump will automatically
restart when the mains power returns.
To remove auto-restart switch off the mains power at the rear of the pump. Hold
down the
STOP
key and turn the mains power switch on. The
!
symbol does not
appear.
Do not use auto-restart for more than 100 starts per
hour. We recommend remote control where a high
number of starts is required.