Stop ramp
This will set the rate of deceleration when the pump stops.
Can be set to a value between 1 and 5.
1 is the fastest deceleration, 5 is the slowest.
Note: Stop ramp is not included in flow calibration.
Delete a recipe
1.
Stop the pump.
2.
From the
CHANGE MODE
menu, Using the
∧
/
∨
keys, scroll to
Dispense
and press
SETTINGS
for
DISPENSE SETTINGS
.
3.
Using the
∧
/
∨
keys, scroll to
Recipes
and press
SELECT
.
4.
Using the
∧
/
∨
keys, scroll to a recipe name and press
SELECT
to edit that recipe. The
EDIT RECIPE
screen is displayed.
5.
Using the
∧
/
∨
keys, scroll to
Delete recipe
and press
SELECT
.
Note:
The pump does not use the recipe name to identify a recipe. The pump uses the numerical
position in the list of recipes to identify a recipe. Deleting a recipe may change that numerical
position. After deleting a recipe, check your batch to make sure the assigned recipe is correct.
Note:
The last remaining recipe cannot be deleted.
Dose adjust
Adjust the recipe volume by ±50 %. Applies adjustment to any active recipe. Set the value to 100% to
use the volume defined in the recipe. Set the value to 150% to use +50% above the volume defined in
the recipe. Set the value to 50% to use -50% below the volume defined in the recipe.
Resume interrupted
When on - the pump will resume a dose if interrupted by power cycle or if the user stops the batch.
The batch will continue from the point it stopped.
When off - the operator must start dosing again following a power cycle. The batch will restart from
the beginning.