12
Care and maintenance
Care and maintenance
Care and maintenance
Care and maintenance
Care and maintenance
The pump is sealed to IP31 and is suitable for wipe down cleaning. Do not use solvents, mechanical
scourers, strong organic acids, or alkali based cleaning solutions.
Remove any tubing, detach the pumphead, and wash the pumphead thoroughly with a mild solution of
detergent in water.
Check moving parts of the rotor from time to time for freedom of movement. Lubricate pivot points and
rollers occasionally with Teflon lubricating oil.
The pump has a good, broad chemical resistance to inorganic acids, saline solutions, alkalis, some
hydrocarbons and a large number of oils and greases. It is suitable for wipe down but not for long term
contact with alcohols. The case may be damaged by contact with strong acids or strong solvents
There are no user serviceable parts inside the pump. The unit should be returned to Watson-Marlow or its
appointed agents or distributors for service.
Error messages
Error messages
Error messages
Error messages
Error messages
If a fault condition is detected the pump will stop. All keys will be disabled. The display will flash the error
number:
Error Error Condition
Suggested Action
0
RAM write error
Attempt to reset by switching power OFF/ON. Otherwise call supplier.
1
RAM corruption
Attempt to reset by switching power OFF/ON. Otherwise call supplier.
2
OTP ROM error/corruption Attempt to reset by switching power OFF/ON. Otherwise call supplier.
3
OTP ROM read error
Attempt to reset by switching power OFF/ON. Otherwise call supplier.
5
Unknown pump type
Check the interface card and cables. Attempt to reset by switching power
OFF/ON. Otherwise call supplier.
7
Display failure
Call supplier.
8
Wrong key press
Attempt keypress again. Attempt to reset by switching power OFF/ON.
9
Motor stalled
Stop pump immediately. Check pumphead and tube. Power OFF/ON may
reset. Otherwise call supplier.
10
Tacho fault
Stop pump immediately. Power OFF/ON may reset. Otherwise call supplier.
14
Overspeed
Stop pump immediately. Power OFF/ON may reset. Otherwise call supplier.
15
Over current
Stop pump immediately. Check system. Power OFF/ON may reset. Otherwise
call supplier.
16
Over voltage
Stop pump immediately. Check mains voltage selector switch. Check supply.
Power OFF/ON may reset. Otherwise call supplier.
17
Under voltage
Stop pump immediately. Check mains voltage selector switch. Check supply.
ON/OFF may reset. Otherwise call supplier.
18
Watchdog error
Attempt to reset by switching power OFF/ON. Otherwise call supplier.
19
Over temperature
Stop pump immediately. Turn OFF. Call supplier.
20
Signal out of range
Check analogue control signal range. Trim signal as required. Otherwise call
supplier.
21
Over signal
Reduce the analogue control signal.
22
No signal
Connect analogue control signal or return to manual control.
25
No network attached
Connect RS232 control cable or return to manual control.
26
RS232 fault
Return to manual control. Check RS232 network. Try again.
27
RS232 lost
Return to manual control. Check RS232 network. Try again.
33
Unrecognised keypress
Attempt keypress again. Attempt to reset by switching power OFF/ON.
Otherwise call supplier.
35
Work overload
Turn OFF. Check power supply. Check pumphead and tubing. Wait 30
minutes. Power ON may reset. Otherwise call supplier.
ERR General error condition
Turn OFF. Call supplier.