Error messages and troubleshooting
71
Detector not
responding to console
Bad or disconnected
cable
Inspect cable
connections, tighten
connectors, or replace
cable.
Configuration problem
Check Ethernet
configuration. For
details, see Empower or
MassLynx online Help.
Leak detected
Leak in detector
Follow the procedure to
resolve a detector leak
sensor error on
Shutter failure message Shutter failed
Power-off and then
power-on the PDA
detector.
Solvent in drain line
Leak from flow cell
gasket
Replace the flow cell
(see
Leak from flow cell
inlet and outlet fittings
Check fittings for
overtightening or
undertightening, and
replace fittings if
necessary.
Status light blinks and
lamp light is off
The detector is running
confidence tests
No corrective action
required. Wait until
tests finish.
Status light blinks and
lamp light is on
Failed startup
diagnostic tests
• Reseat and check
alignment of flow
• Flush the flow cell
(see
).
Dirty flow cell causing
shutter diagnostic test
to fail
Flush the flow cell (see
).
Weak lamp
Replace the lamp (see
).
Detector troubleshooting: (Continued)
Symptom
Possible cause
Corrective action