44
5. Align the leak sensor’s T-bar with the slot in the side of the leak sensor
reservoir, and slide the leak sensor into place.
6. Plug the leak sensor connector into the front of the instrument.
7. In the console, select your detector from the system tree.
8. In the detector information window, click Control > Reset to reset the
detector.
Maintaining the flow cell
The flow cell requires maintenance when
•
the reference spectrum changes;
•
the cell fluid leaks out of the drain tube;
•
the detector cannot initialize but the lamp is in good condition;
•
the detector causes high backpressure.
TP02908
TP02892
Slot in leak
sensor reservoir
T-bar
Leak sensor
installed in
reservoir