715001213, Rev. B
2 of 5
Connecting Your Detector
For detailed installation information and cautionary statements regarding this product, refer to the
following documents:
•
ACQUITY Evaporative Light Sattering Detector Operator’s Guide (71500109303)
•
Ethernet Instrument Getting Started Guide (71500074403)
•
Installing System Leak Sensors (715001557)
•
ACQUITY UPLC Leak Sensor Maintenance Instructions (71500082506). This document is an
addendum to the ACQUITY UPLC System Operator’s Guide.
Electrical Connections
A. Connect any required event or signal cables.
B. Connect an Ethernet cable from the detector to the
system LAN switch or to the
network card on the acquisition computer.
C. Connect the power cable to a grounded receptacle.
B
A
Signal Out +
Signal Out –
Ground
Auxillliary Out +
Auxillliary Out –
Stop Flow +
Stop Flow -
Ground
Switch 2 +
Switch 2 -
1
2
3
4
5
6
7
8
9
10
Inject Start +
Inject Start –
Ground
Lamp On/Off +
Lamp On/Off –
Chart Mark +
Chart Mark –
Ground
Auto Zero +
Auto Zero –
A
B
Ethernet
Cable
441000372
or
441000456
C
Fuse - WAT163-16 (if equipped)
Range: 110 Vac to 240 Vac
Gas Connection
Figure 1 - ACQUITY ELS Rear Electrical and Gas Connections
110 V - WAT097194
220 V - WAT097410
Gas Exhaust