NOTE: 

The following are the basic steps to install the 2465 detector. Check the      after completing each section. For more detailed information and cautionary 

statements regarding  this product, refer to the

 2465

 Electrochmechanical Detector Operator's Manual.

NOTE: 

The following are the basic steps to install the 2465 detector. Check the      after completing each section. For more detailed information and cautionary 

statements regarding  this product, refer to the

 2465

 Electrochemical Detector Operator's Guide.

Electrical Connections

1

Fluid Connections

2

NOTE:

 

 

The following is an example of typical 
fluid connections. Your system may 
require a different setup. All tubing is 
PEEK.

Continued on the back page.

Connectors A and B 

Signals on these connectors let the 
2465 control (or be controlled by) 
external equipment. Together with 
time files, these signals provide a 
powerful tool for the development of 
automated methods.

Refer to the 

2465 Electrochemical 

Detector Operator's Guide 

for more 

information on I/O signal 
connections.

RS-232

Power

Connect an RS-232 cable 
between the detector and 
an Empower™ personal 
computer, workgroup or 
enterprise client.

Connect the power cord to 
a suitable receptacle.

CONNECTOR

WORKING

ELECTRODE

BLOCK

INLET

BLOCK

LC IN

(from column)

LC OUT

(to waste)

AUX

(BLUE)

WORK

(RED)

REF
(BLACK)

FLOW CELL ASSEMBLY

Unpack the unit and lift it from the  
shipping box.

NOTE:

 Never lift the unit by its 

 

front door, only by its sides.

Unpacking 

1.  If using a pulse dampener, run 

tubing from the pulse dampener 
and through the hole at the 
bottom of the side panel. Connect 
this tubing to the inlet (bottom) of 
the column.

2.  Connect the oulet (top) of the 

column to the flow cell's inlet 
block. Rotate the flow cell holder 
to 45º, as shown in the 
illustration, to avoid bubble 
formation. 

 

Also, keep the tubing connecting 
the column to the flow cell as 
short as possible.

3.  Feed a piece of tubing from the 

outlet of the working electrode 
block and through the hole at the 
top of the unit's side panel. This 
tubing goes to waste.

NOTE:

 Place the waste container 

above the level of the flow cell to 
supply back pressure.

Refer to the 

2465 Electrochemical  

Detector Operator's Guide 

for more 

information on fluidic connections.

To install your 2465 Detector:

A

B

RS-232

ANALOG
OUTPUT

1

8

15

9

1

8

15

9

5

9

1

6

For instrument control, 
data acquisition

For chart marker, inject marker.

Event inputs and outputs
(See chart above)

Voltage output via SAT/IN 
to Breeze system, Millenium

32

 

client or workstation, or 
recording device.

GROUND STUD

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

GND

GND

N/C

N/C

MARK

N/C

N/C

N/C

N/C

N/C

N/C

N/C

INJECT MARKER

N/C

N/C

RELAY 1

RELAY 2

CELL ON

RESET

OVERLOAD

AUX 1

AUX 2

CELL OFF

START

AUTO ZERO

GND

Connector A

BLACK

BROWN

RED

ORANGE

YELLOW

GREEN

BLUE

PURPLE

GREY

WHITE

PINK

LIGHT GREEN

BLACK with STRIPE

BROWN with STRIPE

RED with STRIPE

Wire Color

Connector B

IN

OUT

110V  WAT097194

220V  WAT097410