Troubleshooting Noises
84
5
5.2 Troubleshooting Noises
is a guide to troubleshooting and correcting noises in the Waters 1525 Pump.
Table 5-2 Correcting Noises
Symptom
Possible Cause
Corrective Action
Click or loud snap
Binding plunger seal
If the click does not stop and
cannot be isolated to one
pump head, replace the
plunger seals, one at a time
(see
or
).
Worn indicator rod spring
Replace the spring.
Squeak
Plunger seals dry
Moisten the plunger with
appropriate eluent through
pump head access holes.
Binding plunger seal
Replace the plunger seal
assembly (see
).
Improper plunger seal
Install the correct pump
plunger seal (see
or
Binding piston indicator rod
Replace the indicator rod.