Manual for the Electronic Meter "FT 20"

pH                          Temp.                 RH

0.00 - 14.00            0 - 50°C              10 - 99%

Thank you for purchasing our “FT 20" pH - Temperature - RH meter. 
All pH-readings are automatically temperature compensated (ATC). Measurements are 
highly accurate. 

 

FT20 

Range

 

pH: 0.00-14.00 

 

Temp.: 0 -50°C

 

RH: 10 - 99%

Resolution

 

pH: 0.01 pH 

 

Temp.: 0.1°C

 

RH: 1%

Accuracy

 

pH: +/- 0.1 

 

Temp.: +/- 1°C

 

RH: +/- 5%

Calibration

 

manual 

Temp. Comp.

  automatic (ATC)

Battery Type

  4 x 1.5 V 

Batt. Life

 

~ 100 h 

Environment

  0 - 50° C 

 

RH 100% 

pH-Calibration:

1) Remove the protective cap.
2) Switch the meter on by folding back the electrode 
    (automatic switch on/off).
3) Immerse the electrode in a buffer solution “pH 7.00".
5) Stir gently and wait for the reading to stabilize.
6) Use the screw-driver to adjust the reading to “7.00", 
    screwing at the screw marked “STAND“.
7) Rinse the electrode with water.
8) Replace the protective cap and switch the meter off.

Calibrate when:

• The electrode has not been used for a while
• The utmost accuracy is required
• The electrode has been used in particularly taxing conditions

Changing Battery:

1) Open the battery 
    compartment 
    on the back side

2) Remove the used 
    batteries

3) Place new batteries
    4 x 1.5 V (AG-13)
    and use caution to 
    the position of the 
    poles

4) Close the battery 
    compartment

Operation:

1) Remove the protective cap and switch the meter on by 
    folding back the electrode (automatic on/off).
2) Make sure, that the holes on the back side of the meter are 
    not covered during the test procedure.
3) Immerse the electrodes in the solution to be tested.
4) Stir gently and wait for the reading to stabilize.
5) Read the result, than rinse the electrodes with water.
6) Replace the protective cap and switch the meter off.

Manual for the Electronic Meter "FT 20"

pH                          Temp.                 RH

0.00 - 14.00            0 - 50°C              10 - 99%

Thank you for purchasing our “FT 20" pH - Temperature - RH meter. 
All pH-readings are automatically temperature compensated (ATC). Measurements are 
highly accurate. 

 

FT20 

Range

 

pH: 0.00-14.00 

 

Temp.: 0 -50°C

 

RH: 10 - 99%

Resolution

 

pH: 0.01 pH 

 

Temp.: 0.1°C

 

RH: 1%

Accuracy

 

pH: +/- 0.1 

 

Temp.: +/- 1°C

 

RH: +/- 5%

Calibration

 

manual 

Temp. Comp.

  automatic (ATC)

Battery Type

  4 x 1.5 V 

Batt. Life

 

~ 100 h 

Environment

  0 - 50° C 

 

RH 100% 

pH-Calibration:

1) Remove the protective cap.
2) Switch the meter on by folding back the electrode 
    (automatic switch on/off).
3) Immerse the electrode in a buffer solution “pH 7.00".
5) Stir gently and wait for the reading to stabilize.
6) Use the screw-driver to adjust the reading to “7.00", 
    screwing at the screw marked “STAND“.
7) Rinse the electrode with water.
8) Replace the protective cap and switch the meter off.

Calibrate when:

• The electrode has not been used for a while
• The utmost accuracy is required
• The electrode has been used in particularly taxing conditions

Changing Battery:

1) Open the battery 
    compartment 
    on the back side

2) Remove the used 
    batteries

3) Place new batteries
    4 x 1.5 V (AG-13)
    and use caution to 
    the position of the 
    poles

4) Close the battery 
    compartment

Operation:

1) Remove the protective cap and switch the meter on by 
    folding back the electrode (automatic on/off).
2) Make sure, that the holes on the back side of the meter are 
    not covered during the test procedure.
3) Immerse the electrodes in the solution to be tested.
4) Stir gently and wait for the reading to stabilize.
5) Read the result, than rinse the electrodes with water.
6) Replace the protective cap and switch the meter off.