Water I.D. FT 40 Manual Download Page 2

Manual for the Electronic Meter "FT 15"

pH                                         Temp.

0.00 - 14.00                           0 - 55°C

Thank you for purchasing our "FT 15" (splash-water-proof) pH - Temperature 
meter. All pH-readings are automatically temperature compensated (ATC).
Measurements are highly accurate.
The pH electrode, supplied with the meter, can be replaced easily 
by the user.

 

FT15 

Range

 

pH: 0.00-14.00 

 

Temp.: 0 -55°C

Resolution

 

pH: 0.01 pH 

 

Temp.: 0.1°C

Accuracy

 

pH: +/- 0.1 

 

Temp.: +/- 1°C

Calibration

 

manual 

Temp. Comp.

  automatic (ATC)

Battery Type

  4 x 1.5 V 

Batt. Life

 

~ 100 h 

Environment

  0 - 50° C 

 

RH 100%

pH-Calibration:

1) Remove the upper and lower cap.
2) Switch the meter on.
3) Immerse the electrode in a buffer solution "pH 7.00".
4) Stir gently and wait for the reading to stabilize.
5) Use the screw-driver to adjust the reading to "7.00",
    screwing at the screw next to the battery compartment.
6) Rinse the electrode with water.
7) Replace both caps and switch the meter off.

Calibrate when:

• The electrode has not been used for a while
• The utmost accuracy is required
• The electrode has been used in particularly taxing conditions

Changing Battery:

1) Remove the upper 
    cap

2) Remove the used 
    batteries

3) Place new batteries
    4 x 1.5 V (AG-13)
    and use caution to 
    the position of the 
    poles

4) Replace the upper
    cap

Operation:

1) Remove the protective cap and switch the meter on.
2) Immerse the electrode in the solution to be tested.
3) Stir gently and wait for the reading to stabilize.
4) Press “°C/°F“ to switch between Celsius and Fahrenheit.
5) Read the result, than rinse the electrode with water.
6) Replace the protective cap and switch the meter off.

Manual for the Electronic Meter "FT 15"

pH                                         Temp.

0.00 - 14.00                           0 - 55°C

 

FT15 

Range

 

pH: 0.00-14.00 

 

Temp.: 0 -55°C

Resolution

 

pH: 0.01 pH 

 

Temp.: 0.1°C

Accuracy

 

pH: +/- 0.1 

 

Temp.: +/- 1°C

Calibration

 

manual 

Temp. Comp.

  automatic (ATC)

Battery Type

  4 x 1.5 V 

Batt. Life

 

~ 100 h 

Environment

  0 - 50° C 

 

RH 100%

pH-Calibration:

1) Remove the upper and lower cap.
2) Switch the meter on.
3) Immerse the electrode in a buffer solution "pH 7.00".
4) Stir gently and wait for the reading to stabilize.
5) Use the screw-driver to adjust the reading to "7.00",
    screwing at the screw next to the battery compartment.
6) Rinse the electrode with water.
7) Replace both caps and switch the meter off.

Calibrate when:

• The electrode has not been used for a while
• The utmost accuracy is required
• The electrode has been used in particularly taxing conditions

Changing Battery:

1) Remove the upper 
    cap

2) Remove the used 
    batteries

3) Place new batteries
    4 x 1.5 V (AG-13)
    and use caution to 
    the position of the 
    poles

4) Replace the upper
    cap

Operation:

1) Remove the protective cap and switch the meter on.
2) Immerse the electrode in the solution to be tested.
3) Stir gently and wait for the reading to stabilize.
4) Press “°C/°F“ to switch between Celsius and Fahrenheit.
5) Read the result, than rinse the electrode with water.
6) Replace the protective cap and switch the meter off.

Thank you for purchasing our "FT 15" (splash-water-proof) pH - Temperature 
meter. All pH-readings are automatically temperature compensated (ATC).
Measurements are highly accurate.
The pH electrode, supplied with the meter, can be replaced easily 
by the user.