Device Description:

4

Measurement:

1.  After calibration, rinse the electrode with clean 
 

water and wipe it dry. Dip the electrode into 

 

sample solution to be measured. Stir gently 

 

and wait until a stable reading can be obtained.

2.  After measurement, rinse the electrode with 
 

clean water, and replace the soaking bottle and 

 

protective cap. The soaking bottle should be 

 

always kept wet by adding soaking solution.

 

Note:

1.  When doing a 2 points calibration, calibrate 
 

with buffer pH 7 first, then follow with pH 4. 

2.  It is not necessary to calibrate before each 
 

usage. However it should be performed every 

 

two weeks or after 10 times of usage. 

3.  Change a new battery when the display fades 
 

or flashes.