WatchGas Application Note 6 Part 2: WatchGas QGM Calibration v1.1 19-01-21
© 2021 WatchGas B.V.
Automatic calibration using the WatchGas QGM Docking station
To calibrate the WatchGas QGM with Docking Station, use the following steps:
1. Turn on the device and wait for all sensor to warm-up.
2. Turn on the docking station. Ensure the docking station is configured correctly and there’s sufficient
gas remaining on the internal gas cylinder.
3. Place the QGM in the docking station.
4. Press the Calibration button.
5. The docking station will now preform the Zero and Span Calibration automatically.
6. When the Calibration is finished, remove the device from the docking station.